Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, OH, USA.
R1 Engineering LLC, Euclid, OH, USA.
Int J Artif Organs. 2020 Sep;43(9):600-605. doi: 10.1177/0391398820903316. Epub 2020 Feb 4.
This study aimed to evaluate a newly designed circulatory mock loop intended to model cardiac and circulatory hemodynamics for mechanical circulatory support device testing. The mock loop was built with dedicated ports suitable for attaching assist devices in various configurations. This biventricular mock loop uses two pneumatic pumps (Abiomed AB5000, Danvers, MA, USA) driven by a dual-output driver (Thoratec Model 2600, Pleasanton, CA, USA). The drive pressures can be individually modified to simulate a healthy heart and left and/or right heart failure conditions, and variable compliance and fluid volume allow for additional customization. The loop output for a healthy heart was tested at 4.2 L/min with left and right atrial pressures of 1 and 5 mm Hg, respectively; a mean aortic pressure of 93 mm Hg; and pulmonary artery pressure of 17 mm Hg. Under conditions of left heart failure, these values were reduced to 2.1 L/min output, left atrial pressure = 28 mm Hg, right atrial pressure = 3 mm Hg, aortic pressure = 58 mm Hg, and pulmonary artery pressure = 35 mm Hg. Right heart failure resulted in the reverse balance: left atrial pressure = 0 mm Hg, right atrial pressure = 30 mm Hg, aortic pressure = 100 mm Hg, and pulmonary artery pressure = 13 mm Hg with a flow of 3.9 L/min. For biventricular heart failure, flow was decreased to 1.6 L/min, left atrial pressure = 13 mm Hg, right atrial pressure = 13 mm Hg, aortic pressure = 52 mm Hg, and pulmonary artery pressure = 18 mm Hg. This mock loop could become a reliable bench tool to simulate a range of heart failure conditions.
本研究旨在评估一种新设计的循环模拟回路,该回路旨在模拟心脏和循环血液动力学,用于机械循环支持设备测试。该模拟回路采用专用端口构建,可用于各种配置连接辅助设备。这个双心室模拟回路使用两个气动泵(美国亚利桑那州丹佛市的 Abiomed AB5000),由双输出驱动器(美国加利福尼亚州普莱森顿的 Thoratec Model 2600)驱动。驱动压力可以单独调整,以模拟健康心脏和左心或右心衰竭的情况,并且可变性顺应性和液体量允许进行其他定制。健康心脏的回路输出在左心房压力分别为 1 和 5mmHg、平均主动脉压力为 93mmHg 和肺动脉压力为 17mmHg 的情况下测试为 4.2L/min。在左心衰竭的情况下,这些值降低到 2.1L/min 的输出、左心房压力为 28mmHg、右心房压力为 3mmHg、主动脉压力为 58mmHg 和肺动脉压力为 35mmHg。右心衰竭导致相反的平衡:左心房压力为 0mmHg、右心房压力为 30mmHg、主动脉压力为 100mmHg 和肺动脉压力为 13mmHg,流量为 3.9L/min。对于双心室心力衰竭,流量降低至 1.6L/min,左心房压力为 13mmHg、右心房压力为 13mmHg、主动脉压力为 52mmHg 和肺动脉压力为 18mmHg。这个模拟回路可以成为一种可靠的台架工具,用于模拟一系列心力衰竭情况。