Mańka-Malara Katarzyna, Panasiewicz Paulina, Kacprzyk Milena, Gawryszewska Monika, Mierzwińska-Nastalska Elżbieta, Gawlak Dominika
Department of Prosthodontics, Medical University of Warsaw, Warsaw, Poland.
Acta Bioeng Biomech. 2019;21(4):63-71.
Mouthguards are commonly used in the prevention of head and mouth injuries in many sport disciplines. Elastic polymeric materials used in their fabrication absorb the energy of impact. However, decontamination procedures may cause damage to their surface causing deterioration of their properties and favouring pathogen colonization. The aim of the study was to determine the impact of different cleaning methods on mouthguard materials, surface quality.
The material for the research consisted of 81 samples made from polymeric materials used for custom mouthguard fabrication - Impak, Corflex Orthodontic and Erkoflex. Samples were viewed under a scanning electron microscope after exposure to different cleaning methods.
Statistical analysis of tested decontamination procedures and disinfecting agents showed statistically important ( p < 0.05) results for the presence of small and big separated pieces, grooves and holes. Out of all the cases the lowest indicator for occurrence was in the disinfecting spray - Aftermat.
Considering the importance of the surface material of the mouthguard, disinfection should be the method of choice for maintaining its hygiene. Impak material samples, cleaned with the Aftermat spray, showed the lowest range of damage.
护齿器在许多体育项目中常用于预防头部和口腔损伤。用于制造护齿器的弹性聚合物材料可吸收撞击能量。然而,去污程序可能会损坏其表面,导致其性能下降并有利于病原体定植。本研究的目的是确定不同清洁方法对护齿器材料表面质量的影响。
研究材料包括81个由用于定制护齿器制造的聚合物材料制成的样品——Impak、Corflex Orthodontic和Erkoflex。在暴露于不同清洁方法后,在扫描电子显微镜下观察样品。
对测试的去污程序和消毒剂的统计分析表明,对于小的和大的分离碎片、凹槽和孔洞的存在,结果具有统计学意义(p < 0.05)。在所有情况中,发生率最低的指标出现在消毒喷雾——Aftermat中。
考虑到护齿器表面材料的重要性,消毒应是保持其卫生的首选方法。用Aftermat喷雾清洁的Impak材料样品显示出最低的损伤范围。