Multicenter Evaluation of the Quantra QPlus System in Adult Patients Undergoing Major Surgical Procedures.

机构信息

From the Department of Anesthesiology, University of Virginia, Charlottesville, Virginia.

Department of Anesthesiology, Duke University, Durham, North Carolina.

出版信息

Anesth Analg. 2020 Apr;130(4):899-909. doi: 10.1213/ANE.0000000000004659.

Abstract

BACKGROUND

The management of perioperative bleeding and the optimization of the available therapies are subjects of significant clinical interest. Clinical guidelines recommend the use of whole blood viscoelastic testing devices to target the utilization of blood products during major surgical procedures. The Quantra QPlus System is a new cartridge-based viscoelastic testing device based on an innovative ultrasound technology. The aim of this study was to evaluate this new system in a surgical population.

METHODS

Two hundred seventy-seven adult subjects were enrolled in a multicenter, prospective observational study consisting primarily of patients undergoing cardiac and major orthopedic surgeries. Samples were obtained at multiple time points for testing on the Quantra QPlus System, the rotational thromboelastometry (ROTEM) delta, and standard coagulation tests. Quantra measurements included Clot Time (CT), Heparinase Clot Time (CTH), Clot Time Ratio (CTR), Clot Stiffness (CS), Fibrinogen (FCS), and Platelet (PCS) Contributions to CS. Data analyses included assessment of the concordance of Quantra parameters with a series of clinical composite indexes formed on the basis of standard coagulation tests in 3 domains representing increased, decreased, and normal/subclinical coagulation function. Linear regression and receiver operator characteristic (ROC) analyses of Quantra parameters with corresponding parameters from ROTEM assays were also performed.

RESULTS

The accuracy (overall percent agreement or ratio of true positives and true negatives over the entire population) between the Quantra and the composite indexes was between 72% and 98% depending on the specific parameter. Linear regression analysis indicated that the correlation between ROTEM delta and Quantra was very strong with r values ranging between 0.84 and 0.89. Results from ROC analysis demonstrated sensitivities and specificities in the 80%-90% range when QPlus parameters were used to discriminate ROTEM threshold values currently used in goal-directed treatment algorithms.

CONCLUSIONS

This study demonstrated that the Quantra QPlus System is strongly correlated with a well-established viscoelastic testing device and its parameters effectively represent the results from multiple standard laboratory assays. The Quantra has been designed to operate at the point of care with the potential to provide rapid and comprehensive results to aid in the management of coagulopathic patients.

摘要

背景

围手术期出血的管理和可用治疗方法的优化是具有重要临床意义的课题。临床指南建议使用全血粘弹性测试设备,以针对主要手术过程中血液制品的使用。Quantra QPlus 系统是一种基于创新超声技术的新型基于卡匣的粘弹性测试设备。本研究旨在评估该新系统在手术人群中的应用。

方法

277 名成年患者参与了一项多中心前瞻性观察研究,主要包括接受心脏和主要骨科手术的患者。在多个时间点采集样本,用于 Quantra QPlus 系统、旋转血栓弹性测定仪(ROTEM) delta 和标准凝血测试。Quantra 测量包括凝血时间(CT)、肝素酶凝血时间(CTH)、凝血时间比(CTR)、凝血硬度(CS)、纤维蛋白原(FCS)和血小板(PCS)对 CS 的贡献。数据分析包括评估 Quantra 参数与一系列临床综合指数的一致性,这些指数是基于 3 个代表凝血功能增加、降低和正常/亚临床的标准凝血测试领域形成的。还对 Quantra 参数与 ROTEM 测定相应参数之间进行了线性回归和接收者操作特征(ROC)分析。

结果

根据特定参数,Quantra 与复合指数之间的准确性(整个人群中真实阳性和真实阴性的总百分比或比率)在 72%至 98%之间。线性回归分析表明,ROTEM delta 和 Quantra 之间的相关性非常强,r 值在 0.84 至 0.89 之间。ROC 分析结果表明,当使用 QPlus 参数来区分目前用于目标导向治疗算法的 ROTEM 阈值时,参数的敏感性和特异性在 80%-90%范围内。

结论

本研究表明,Quantra QPlus 系统与一种成熟的粘弹性测试设备具有很强的相关性,其参数有效地代表了多个标准实验室检测的结果。Quantra 的设计目的是在护理点运行,有可能提供快速全面的结果,以帮助管理凝血功能障碍患者。