心脏手术中 SEER 声学法鼻咽喉测量与标准凝血试验、旋转血栓弹性描记术和多电极聚集法的比较研究。

A Comparative Study of SEER Sonorheometry Versus Standard Coagulation Tests, Rotational Thromboelastometry, and Multiple Electrode Aggregometry in Cardiac Surgery.

机构信息

Department of Cardiovascular Anaesthesia and Intensive Care Unit, IRCCS Policlinico San Donato, San Donato Milanese (Milan), Milan, Italy.

Department of Cardiovascular Anaesthesia and Intensive Care Unit, IRCCS Policlinico San Donato, San Donato Milanese (Milan), Milan, Italy.

出版信息

J Cardiothorac Vasc Anesth. 2019 Jun;33(6):1590-1598. doi: 10.1053/j.jvca.2019.01.011. Epub 2019 Jan 4.

Abstract

OBJECTIVES

Coagulation point-of-care tests for the diagnosis and management of bleeding commonly are used in cardiac surgery. Recently, a new viscoelastic point-of-care device, Quantra (HemoSonics, Charlottesville, VA), which is based on sonic estimation of elasticity via resonance sonorheometry, entered the market. Prior studies have compared the Quantra parameters with conventional coagulation tests and thromboelastography or rotational thromboelastography, but no study has assessed the correlation of the platelet-derived Quantra parameter platelet contribution to stiffness (PCS) with platelet function assays. The present study compared the Quantra-derived coagulation parameters with conventional coagulation tests, rotational thromboelastography-derived parameters, and platelet function measured using multiplate aggregometry.

DESIGN

Prospective cohort study.

SETTING

University research hospital.

PARTICIPANTS

The study comprised 30 cardiac surgery patients before and after cardiopulmonary bypass.

INTERVENTIONS

None.

MEASUREMENTS AND MAIN RESULTS

Quantra clot stiffness showed a very strong correlation with the EXTEM maximum clot firmness before and after cardiopulmonary bypass and in the overall samples (r values 0.94-0.96). Quantra fibrinogen contribution to stiffness was compared with the corresponding FIBTEM and Clauss fibrinogen levels, which also showed a strong level of correlation (r values between 0.74 and 0.87). Quantra PCS strongly correlated with platelet count (r = 0.71) in the overall samples and moderately with adenosine diphosphate-dependent platelet function (r = 0.67). In a multivariable model, both the adenosine diphosphate test value and the platelet count remained independently associated with Quantra PCS.

CONCLUSIONS

Fibrinogen-dependent clot stiffness properties are well-reflected by the Quantra fibrinogen contribution to stiffness parameter, and PCS incorporates platelet count and function.

摘要

目的

用于诊断和管理出血的即时凝血检测点在心脏手术中通常被使用。最近,一种新的动态即时凝血检测设备 Quantra(HemoSonics,弗吉尼亚州夏洛茨维尔),基于共振声学法通过对弹性的声波估计,进入市场。先前的研究比较了 Quantra 参数与常规凝血测试和血栓弹力描记法或旋转血栓弹力描记法,但没有研究评估血小板衍生的 Quantra 参数血小板对硬度的贡献(PCS)与血小板功能测定的相关性。本研究比较了 Quantra 衍生的凝血参数与常规凝血测试、旋转血栓弹力描记法衍生的参数以及使用多板聚集仪测量的血小板功能。

设计

前瞻性队列研究。

设置

大学研究医院。

参与者

该研究包括 30 名心脏手术患者,在体外循环前后。

干预措施

无。

测量和主要结果

Quantra 凝块硬度与体外循环前后 EXTEM 最大凝块硬度以及总样本中 EXTEM 最大凝块硬度非常强相关(r 值为 0.94-0.96)。 Quantra 纤维蛋白原对硬度的贡献与对应的 FIBTEM 和 Clauss 纤维蛋白原水平进行了比较,也显示出很强的相关性(r 值在 0.74 和 0.87 之间)。 Quantra PCS 与血小板计数在总样本中呈强相关(r 值为 0.71),与二磷酸腺苷依赖性血小板功能中度相关(r 值为 0.67)。在多变量模型中,二磷酸腺苷测试值和血小板计数都与 Quantra PCS 独立相关。

结论

Quantra 纤维蛋白原对硬度的贡献参数很好地反映了纤维蛋白原依赖性凝块硬度特性,而 PCS 包含血小板计数和功能。