Apheresis and Cellular Therapy Unit, ICNU, Department of Hemotherapy and Hemostasis, ICMHO, IDIBAPS, Hospital Clínic, University of Barcelona, Barcelona, Spain.
Transfusion. 2020 Mar;60(3):607-612. doi: 10.1111/trf.15698. Epub 2020 Feb 10.
Peripheral venous access (PVA) devices for apheresis should be selected to provide enough flow with the least vein damage, but little information is available about PVA devices blood flow rates. The aim of this study was to know the blood flow rates provided by the most frequent PVA devices used in apheresis procedures.
An experimental study was designed to simulate a plasma exchange procedure using citrated whole blood (WB). Two steel needles (16- and 17-gauge [G]) and six plastic cannulas (16G, 17G, two 18G, 20G, and 22G) were analyzed. The hematocrit of the WB was adjusted to 45%, 40%, 35%, 30%, and 25%. The separated plasma from the WB was used as replacement fluid.
Blood flow rate (inlet/return, mL/min) for 16G and 17G devices with a hematocrit of 45% was 142/142 (maximum admitted by separator); one of the 18G cannulas reached 142/142 and the other one reached 117/140; the 20G cannula reached 78/94; and the 22G reached 45/55. A hematocrit reduction from 45% to 25% increased the flow rate (when possible) over 22% (range: 14%-30%).
PVA devices with a size of 16-18G provided the maximum flow admitted by the apheresis system. The 20G provided flow rates for a significant number of procedures, and 22G could be used in some procedures. The hematocrit should be taken into account when selecting the PVA device since, at hematocrit of 25%, the flow can be from 14% to 30% higher than flow rates reached with blood at hematocrit of 45%.
用于单采的外周静脉通路(PVA)装置应选择提供足够的血流量,同时对血管损伤最小,但关于 PVA 装置血流量的信息很少。本研究旨在了解单采过程中最常使用的 PVA 装置提供的血流量。
设计了一项实验研究,使用柠檬酸盐全血(WB)模拟血浆置换程序。分析了两根钢针(16 号和 17 号)和六根塑料套管(16G、17G、两个 18G、20G 和 22G)。WB 的血细胞比容调整为 45%、40%、35%、30%和 25%。从 WB 中分离的血浆用作替代液。
血细胞比容为 45%的 16G 和 17G 装置的血流速度(进口/返回,mL/min)为 142/142(分离器允许的最大值);其中一个 18G 套管达到 142/142,另一个达到 117/140;20G 套管达到 78/94;22G 达到 45/55。血细胞比容从 45%降至 25%时,(在可能的情况下)将流速提高了 22%以上(范围:14%-30%)。
尺寸为 16-18G 的 PVA 装置提供了单采系统允许的最大流量。20G 提供了大量程序的流量,22G 可用于某些程序。在选择 PVA 装置时应考虑血细胞比容,因为在血细胞比容为 25%时,与血细胞比容为 45%时达到的流量相比,流量可提高 14%至 30%。