Abstract

We investigated the performance of the qCON index regarding its time delay for sudden changes in the anesthetic level as well as to separate responsiveness from unresponsiveness during loss and return of responsiveness (LOR and ROR). For evaluation of the time delay, we replayed relevant EEG episodes to the qCON to simulate sudden changes between the states (i) awake/sedation, (ii) adequate anesthesia, or (iii) suppression. We also replayed EEG from 40 patients during LOR and ROR to evaluate the qCON's ability to separate responsiveness from unresponsiveness. The time delays depended on the type of transition. The delays for the important transition between awake/sedation and adequate anesthesia were 21(5) s from awake/sedation to adequate anesthesia and 26(5) s in the other direction. The performance of the qCON to separate responsiveness from unresponsiveness depended on signal quality, the investigation window, i.e. ± 30 s or ± 60 s around LOR/ROR, and the specific transition being tested. AUC was 0.63-0.90 for LOR and 0.61-0.79 for ROR. Time delay and performance during state transitions of the qCON were similar to other monitoring systems such as bispectral index. The better performance of qCON during LOR than ROR probably reflects the sudden change in EEG activity during LOR and the more heterogeneous EEG during ROR.

摘要

我们研究了 qCON 指数在麻醉水平突然变化时的时滞性能,以及在失去和恢复反应性(LOR 和 ROR)期间区分反应性和无反应性。为了评估时滞,我们将相关的 EEG 片段重放到 qCON 中,以模拟状态之间的突然变化(i)清醒/镇静,(ii)适当的麻醉,或(iii)抑制。我们还重放了 40 名患者在 LOR 和 ROR 期间的 EEG,以评估 qCON 区分反应性和无反应性的能力。时滞取决于过渡类型。从清醒/镇静到适当麻醉的重要过渡的时滞为 21(5) s,而在另一个方向上为 26(5) s。qCON 区分反应性和无反应性的性能取决于信号质量、调查窗口,即 LOR/ROR 周围的±30 s 或±60 s,以及正在测试的特定过渡。LOR 的 AUC 为 0.63-0.90,ROR 的 AUC 为 0.61-0.79。qCON 在状态过渡期间的时滞和性能与其他监测系统(如双谱指数)相似。qCON 在 LOR 期间的性能优于 ROR,可能反映了 LOR 期间 EEG 活动的突然变化,以及 ROR 期间 EEG 的异质性更大。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e519/7943427/b07a20049ce8/10877_2020_480_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验