Haase A, Leibfritz D, Werk W
Universität Bremen, Federal Republic of Germany.
Magn Reson Med. 1988 Jul;7(3):358-63. doi: 10.1002/mrm.1910070314.
A method of 31P FLASH NMR imaging is described using a low flip angle excitation pulse and three-dimensional Fourier transform spectroscopic imaging (3DFT). In vivo high-resolution 31P NMR spectra are obtained for each of 32 x 32 image elements using a minimum measuring time of 1.5 min. 31P FLASH images of a rat showing the spatial distribution of phosphocreatine, alpha-, beta-, and gamma-ATP have been measured in a 4.7-T magnet with a spatial resolution of 6 mm.
描述了一种使用低翻转角激发脉冲和三维傅里叶变换光谱成像(3DFT)的31P FLASH NMR成像方法。使用1.5分钟的最短测量时间,为32×32个图像元素中的每一个获取体内高分辨率31P NMR光谱。在4.7-T磁体中测量了一只大鼠的31P FLASH图像,显示了磷酸肌酸、α-、β-和γ-ATP的空间分布,空间分辨率为6毫米。