Gao H M, Song S Z, Mei Y, Nie M H, Fang R D, Su W T
School of Public Health, Wuhan University of Science and Technology, Hubei Province Key Laboratory of Occupational Hazard Identification and Control, Wuhan 430065, China.
Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi. 2020 Jan 20;38(1):52-55. doi: 10.3760/cma.j.issn.1001-9391.2020.01.013.
To establish a method for the determination of 1-methoxy-2-propanol in urine using headspace solid phase micro-extraction coupled with gas chromatography. The 1-methoxy-2-propanol was enriched by headspace solid phase micro-extraction fiber coated with carbene/polydimethylsiloxane (CAR/PDMS) . Single factor rotation method was used to optimize the conditions of extraction temperature, salt amount, and extraction time. The separation was performed on DB-5 (30 m×0.32 mm×0.25 μm) capillary column and detected with flame ionization detector. The quantification was based on the standard curve. The concentration of 1-methoxy-2-propanol in urine was linear in the range of 0.50-10.0 mg/L, and the linear correlation coefficient was 0.9993. The detection limit of the method was 0.14 mg/L, and the limit of quantification was 0.45 mg/L. The recovery was 85.8% to 104.7%, and the of intra- and inter-batch precision were 3.25%-6.65% and 0.81%-3.96%, respectively. The method is high sensitivity and simple operation, and is suitable for the determination of 1-methoxy-2-propanol in urine of occupational exposure population.