Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, FL, USA.
Department of Ophthalmology, Faculty of Medicine, Benha University, Benha, Egypt.
Curr Eye Res. 2020 Jun;45(6):659-667. doi: 10.1080/02713683.2019.1691238. Epub 2020 Feb 16.
: To characterize the three-dimensional (3D) thickness profile and age-related changes of Bowman's layer (BL), and endothelium/Descemet's membrane (En/DM) complex among healthy individuals using Corneal Microlayer Tomography (CML-T), and to describe its repeatability and accuracy.: Sixty-six eyes of 41 healthy volunteers; 27 eyes (< 40 years old), and 39 eyes (>40 years old) were imaged using HD-OCT. Automatic and manual segmentation of the corneal layers was performed, and 3D thickness maps were generated, using custom-built CML-T software. A regional analysis of mean thickness parameters between the 2 age groups was performed. A regression analysis was used to assess the correlation between age, and thickness maps. Intraclass Correlation Coefficients (ICC), Coefficients of Variation (COV), and Bland-Altman plots were used to assess the reliability of the repeated measurements in 198 locations.: CML-T successfully mapped the BL and En/DM in all included eyes. Thickness maps showed a significant increase in corneal thickness (CT), BL thickness (BMT), and En/DM complex thickness (DMT) toward the periphery with a mean difference 28 µm ( < .001), 1.1 µm ( < .001), and 1.4 µm ( < .001), respectively. There was a strong correlation between age and central DMT (r = 0.61; < .001), while there was no correlation between age and both CT, and BMT. ICC values ranged from 0.9 (BMT) to 0.997 (DMT), and from 0.808 (BMT) to 0.979 (CT) for intraoperator repeatability of manual measurements, and the accuracy of auto matic measurements, respectively. COV values were lower than 7.5% in all cases.: CML-T is a novel tool that can generate 3D-thickness maps of both BL and En/DM. CT, BMT, and DMT increase toward the periphery in healthy corneas. DMT increases with aging, while BMT does not. We also report excellent repeatability, accuracy and good agreement between automatic and manual measurements.
: 利用角膜微层层析成像术(CML-T)描述健康个体的 Bowman 层(BL)和内皮/Descemet 膜(En/DM)复合体的三维(3D)厚度轮廓和与年龄相关的变化,并描述其重复性和准确性。: 对 41 名健康志愿者的 66 只眼进行成像;27 只眼(<40 岁)和 39 只眼(>40 岁)。使用高清-OCT 对角膜进行自动和手动分层,并使用定制的 CML-T 软件生成 3D 厚度图。对 2 个年龄组之间的平均厚度参数进行区域分析。回归分析用于评估年龄与厚度图之间的相关性。使用组内相关系数(ICC)、变异系数(COV)和 Bland-Altman 图评估 198 个位置重复测量的可靠性。: CML-T 成功地绘制了所有纳入眼的 BL 和 En/DM。厚度图显示角膜厚度(CT)、BL 厚度(BMT)和 En/DM 复合体厚度(DMT)随周边逐渐增加,平均差异分别为 28µm(<.001)、1.1µm(<.001)和 1.4µm(<.001)。年龄与中央 DMT 之间存在很强的相关性(r=0.61;<.001),而年龄与 CT 和 BMT 之间均无相关性。内操作员手动测量的重复性的 ICC 值范围为 0.9(BMT)至 0.997(DMT),自动测量的准确性的 ICC 值范围为 0.808(BMT)至 0.979(CT)。在所有情况下,COV 值均低于 7.5%。: CML-T 是一种新颖的工具,可以生成 BL 和 En/DM 的 3D 厚度图。健康角膜的 CT、BMT 和 DMT 随周边逐渐增加。DMT 随年龄增长而增加,而 BMT 则不会。我们还报告了出色的重复性、准确性以及自动和手动测量之间的良好一致性。