Center of Digital Dentistry, Faculty of Prosthodontics, Peking University School and Hospital of Stomatology & National Engineering Laboratory for Digital and Material Technology of Stomatology & Research Center of Engineering and Technology for Digital Dentistry of Ministry of Health & Beijing Key Laboratory of Digital Stomatology & National Clinical Research Center for Oral Diseases, Beijing 100081, China.

出版信息

Beijing Da Xue Xue Bao Yi Xue Ban. 2020 Feb 18;52(1):129-137. doi: 10.19723/j.issn.1671-167X.2020.01.021.

Abstract

OBJECTIVE

To provide a reference for using intraoral scanners for making clinical diagnostic dentures of edentulous jaws by comparing the accuracy of three intraoral scanners for primary impression and jaw relation record of edentulous jaws.

METHODS

This study contained 6 primary impressions of the edentulous patients. Each of the impressions consisted of the maxillary primary impression, the mandibular primary impression and the jaw relation record. For each of them, a dental cast scanner (Dentscan Y500) was used to obtain stereolithography (STL) data as reference scan, and then three intraoral scanners including i500, Trios 3 and CEREC Primescan were used for three times to obtain STL data as experiment groups. In Geomagic Studio 2013 software, trueness was obtained by comparing experiment groups with the reference scan, and the precision was obtained from intragroup comparisons. Registered maxillary data of the intraoral scan with reference scan, the morphological error of jaw relation record was obtained by comparing jaw relation record of the intraoral scan with the reference scan. Registered mandibular data with jaw relation record of intraoral scan and the displacement of the jaw position were evaluated. Independent samples t test and Mann-Whitney U test in the SPSS 20.0 statistical software were used to statistically analyze the trueness, precision and morphological error of jaw relation record of three intraoral scanners. The Bland-Altman diagram was used to evaluate the consistency of the jaw relationship measured by the three intraoral scanners.

RESULTS

The trueness of i500, Trios 3 and CEREC Primescan scanners was (182.34±101.21) μm, (145.21±71.73) μm, and (78.34±34.79) μm for maxilla; (106.42±21.63) μm, and 95.08 (63.08) μm, (78.45±42.77) μm for mandible. There was no significant difference in trueness of the three scanners when scanning the maxilla and mandible(P>0.05). The precision of the three scanners was 147.65 (156.30) μm, (147.54±83.33) μm, and 40.30 (32.80) μm for maxilla; (90.96±30.77) μm, (53.73±23.56) μm, and 37.60 (93.93) μm for mandible. The precision of CEREC Primescan scanner was significantly better than that of the other two scanners for maxilla (P<0.05). Trios 3 and CEREC Primescan scanners were significantly better than i500 scanner for mandible (P<0.05). The precision of the i500 and Trios 3 scanners for mandible was superior to maxilla (P<0.05). The upper limit of 95% confidence intervals of trueness and precision of three scanners for both maxilla and mandible were within ±300 μm which was clinically accepted. The morphological error of jaw relation record of the three scanners was (337.68±128.54) μm, (342.89±195.41) μm, and (168.62±88.35) μm. The 95% confidence intervals of i500 and Trios 3 scanners were over 300 μm. CEREC Primescan scanner was significantly superior to i500 scanner(P<0.05).The displacement of the jaw position of the three scanners was (0.83±0.56) mm, (0.80±0.45) mm, and (0.91±0.75) mm for vertical dimension; (0.79±0.58) mm, (0.62±0.18) mm, and (0.53±0.53) mm for anterior and posterior directions; (0.95±0.59) mm, (0.69±0.45) mm, and (0.60±0.22) mm for left and right directions. The displacement of the jaw position of the three scanners in vertical dimension, anterior and posterior directions and the left and right directions were within the 95% consistency limit.

CONCLUSION

Three intraoral scanners showed good trueness and precision. The i500 and Trios 3 scanners had more errors in jaw relation record, but they were used as primary jaw relation record. It is suggested that three intraoral scanners can be used for obtaining digital data to make diagnostic dentures and individual trays, reducing possible deforming or crack when sending impressions from clinic to laboratory.

摘要

目的

通过比较三种口内扫描仪在无牙颌初印模及颌位关系记录方面的准确性,为使用口内扫描仪制作无牙颌临床诊断义齿提供参考。

方法

本研究纳入6例无牙颌患者的初印模。每个印模包括上颌初印模、下颌初印模及颌位关系记录。对于每个印模,使用牙科模型扫描仪(Dentscan Y500)获取立体光刻(STL)数据作为参考扫描,然后使用包括i500、Trios 3和CEREC Primescan在内的三种口内扫描仪进行三次扫描以获取STL数据作为实验组。在Geomagic Studio 2013软件中,通过将实验组与参考扫描进行比较获得准确性,并从组内比较中获得精度。将口内扫描的上颌数据与参考扫描进行配准,通过将口内扫描的颌位关系记录与参考扫描进行比较获得颌位关系记录的形态误差。将下颌数据与口内扫描的颌位关系记录进行配准并评估颌位的位移。使用SPSS 20.0统计软件中的独立样本t检验和Mann-Whitney U检验对三种口内扫描仪的准确性、精度及颌位关系记录的形态误差进行统计学分析。使用Bland-Altman图评估三种口内扫描仪测量的颌关系的一致性。

结果

i500、Trios 3和CEREC Primescan扫描仪在上颌的准确性分别为(182.34±101.21)μm、(145.21±71.73)μm和(78.34±34.79)μm;在下颌分别为(106.42±21.63)μm、95.08(63.08)μm和(78.45±42.77)μm。三种扫描仪在上颌和下颌扫描时的准确性无显著差异(P>0.05)。三种扫描仪在上颌的精度分别为147.65(156.30)μm、(147.54±83.33)μm和40.30(32.80)μm;在下颌分别为(90.96±30.77)μm、(53.73±23.56)μm和37.60(93.93)μm。CEREC Primescan扫描仪在上颌的精度显著优于其他两种扫描仪(P<0.05)。Trios 3和CEREC Primescan扫描仪在下颌的精度显著优于i500扫描仪(P<0.05)。i500和Trios 3扫描仪在下颌的精度优于上颌(P<0.05)。三种扫描仪在上颌和下颌的准确性及精度的95%置信区间上限均在临床可接受的±300μm范围内。三种扫描仪颌位关系记录的形态误差分别为(337.68±128.54)μm、(342.89±195.41)μm和(168.62±88.35)μm。i500和Trios 3扫描仪的95%置信区间超过300μm。CEREC Primescan扫描仪显著优于i500扫描仪(P<0.05)。三种扫描仪在垂直方向、前后方向和左右方向的颌位位移分别为(0.83±0.56)mm、(0.80±0.45)mm和(0.91±0.75)mm;(0.79±0.58)mm、(0.62±0.18)mm和(0.53±0.53)mm;(0.95±0.59)mm、(0.69±0.45)mm和(0.60±0.22)mm。三种扫描仪在垂直方向、前后方向和左右方向的颌位位移均在95%一致性限度内。

结论

三种口内扫描仪显示出良好的准确性和精度。i500和Trios 3扫描仪在颌位关系记录方面存在较多误差,但可作为颌位关系的初步记录。建议三种口内扫描仪可用于获取数字数据以制作诊断义齿和个别托盘,减少从临床向实验室传送印模时可能出现的变形或破裂。

相似文献

1
[Accuracy of three intraoral scans for primary impressions of edentulous jaws].[三种口腔内扫描用于无牙颌初印模的准确性]
Beijing Da Xue Xue Bao Yi Xue Ban. 2020 Feb 18;52(1):129-137. doi: 10.19723/j.issn.1671-167X.2020.01.021.

本文引用的文献

1
[Research and development of digital design and fabrication of complete denture].全口义齿数字化设计与制造的研究与开发
Zhonghua Kou Qiang Yi Xue Za Zhi. 2018 Jan 9;53(1):60-65. doi: 10.3760/cma.j.issn.1002-0098.2018.01.013.
8
Accuracy of five intraoral scanners compared to indirect digitalization.与间接数字化相比,五种口腔内扫描仪的准确性。