Suppr超能文献

[可穿戴生理监测系统在不同活动条件下进行心肺生理测量的准确性研究]

[Study on the accuracy of cardiopulmonary physiological measurements by a wearable physiological monitoring system under different activity conditions].

作者信息

Xu Haoran, Chu Wenya, Liu Xiaoli, Zhang Shasha, Yang Zhicheng, Zheng Jiewen, Gao Xiaolin, Zhang Zhengbo, Cao Desen

机构信息

Medical School of Chinese PLA, Beijing 100853, P.R.China.

Beijing Health Regulation Co.,Ltd, Beijing 100853, P.R.China.

出版信息

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2020 Feb 25;37(1):119-128. doi: 10.7507/1001-5515.201903061.

Abstract

This paper aims to study the accuracy of cardiopulmonary physiological parameters measurement under different exercise intensity in the accompanying (wearable) physiological parameter monitoring system. SensEcho, an accompanying physiological parameter monitoring system, and CORTEX METALYZER 3B, a cardiopulmonary function testing system, were used to simultaneously collect the cardiopulmonary physiological parameters of 28 healthy volunteers (17 males and 11 females) in various exercise states, such as standing, lying down and Bruce treadmill exercise. Bland-Altman analysis, correlation analysis and other methods, from the perspective of group and individual, were used to contrast and analyze the two types of equipment to measure parameters of heart rate and breathing rate. The results of group analysis showed that the heart rate and respiratory rate data box charts collected by the two devices were highly consistent. The heart rate difference was (-0.407 ± 3.380) times/min, and the respiratory rate difference was (-0.560 ± 7.047) times/min. The difference was very small. The Bland-Altman plot of the heart rate and respiratory rate in each experimental stage showed that the proportion of mean ± 2SD was 96.86% and 95.29%, respectively. The results of individual analysis showed that the correlation coefficients of the whole-process heart rate and respiratory rate data were all greater than 0.9. In conclusion, SensEcho, as an accompanying physiological parameter monitoring system, can accurately measure the human heart rate, respiration rate and other key cardiopulmonary physiological parameters under various sports conditions. It can maintain good stability under various sports conditions and meet the requirements of continuous physiological signal collection and analysis application under sports conditions.

摘要

本文旨在研究在伴随(可穿戴)生理参数监测系统中,不同运动强度下心肺生理参数测量的准确性。使用伴随生理参数监测系统SensEcho和心肺功能测试系统CORTEX METALYZER 3B,同时采集28名健康志愿者(17名男性和11名女性)在站立、躺卧和布鲁斯跑步机运动等各种运动状态下的心肺生理参数。采用布兰德-奥特曼分析、相关性分析等方法,从组和个体的角度,对这两种设备测量心率和呼吸频率的参数进行对比分析。组分析结果表明,两种设备采集的心率和呼吸频率数据箱线图高度一致。心率差值为(-0.407±3.380)次/分钟,呼吸频率差值为(-0.560±7.047)次/分钟。差异非常小。各实验阶段心率和呼吸频率的布兰德-奥特曼图显示,均值±2SD的比例分别为96.86%和95.29%。个体分析结果表明,全过程心率和呼吸频率数据的相关系数均大于0.9。综上所述,SensEcho作为一种伴随生理参数监测系统,能够在各种运动条件下准确测量人体心率、呼吸频率等关键心肺生理参数。它在各种运动条件下都能保持良好的稳定性,满足运动条件下连续生理信号采集与分析应用的要求。

相似文献

本文引用的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验