TLD-100在适用于大分割放疗的剂量范围内的剂量反应。

Dose-Response of TLD-100 in the Dose Range Useful for Hypofractionated Radiotherapy.

作者信息

Liuzzi Raffaele, Piccolo Consiglia, D'Avino Vittoria, Clemente Stefania, Oliviero Caterina, Cella Laura, Pugliese Mariagabriella

机构信息

Institute of Biostructure and Bioimaging, National Research Council (CNR), Naples, Italy.

Department of Physics, State University of Milan, Milan, Italy.

出版信息

Dose Response. 2020 Feb 13;18(1):1559325819894081. doi: 10.1177/1559325819894081. eCollection 2020 Jan-Mar.

Abstract

PURPOSE

The aim of the study was to exploit the feasibility of thermoluminescent dosimeters (TLDs) in radiation therapy techniques in which high dose per fraction is involved.

METHODS

Dose-response of TLD-100 (LiF: Mg, Ti) was investigated in both 6-MV photon and 6-MeV electron beams. The element correction factor (ECF) generation method was applied to check the variability of the TLDs response. Two batches of 50 TLDs were divided into groups and exposed in the dose range 0 to 30 Gy. Regression analysis was performed with both linear and quadratic models. For each irradiation beam, the calibration curves were obtained in 3 dose range 0 to 8 Gy, 0 to 10 Gy, and 0 to 30 Gy. The best-fitting model was assessed by the Akaike Information Criterion test.

RESULTS

The ECF process resulted a useful tool to reduce the coefficients of variation from original values higher than 5% to about 3.5%, for all the batches exposed. The results confirm the linearity of dose-response curve below the dose level of 10 Gy for photon and electron beam and the supralinear trend above.

CONCLUSION

The TLDs are suitable dosimeters for dose monitoring and verification in radiation treatment involving dose up to 30 Gy in a single fraction.

摘要

目的

本研究旨在探讨热释光剂量计(TLD)在涉及高分次剂量的放射治疗技术中的可行性。

方法

研究了TLD-100(LiF:Mg,Ti)在6兆伏光子束和6兆伏电子束中的剂量响应。应用元素校正因子(ECF)生成方法来检查TLD响应的变异性。将两批各50个TLD分成若干组,在0至30戈瑞的剂量范围内进行照射。使用线性和二次模型进行回归分析。对于每种照射束,在0至8戈瑞、0至10戈瑞和0至30戈瑞这3个剂量范围内获得校准曲线。通过赤池信息准则检验评估最佳拟合模型。

结果

对于所有照射批次,ECF方法成为一种有用的工具,可将变异系数从高于5%的原始值降低至约3.5%。结果证实,对于光子束和电子束,在剂量水平低于10戈瑞时剂量响应曲线呈线性,高于该剂量水平时呈超线性趋势。

结论

TLD是适用于单次剂量高达30戈瑞的放射治疗中剂量监测和验证的剂量计。