Seshadri Dhruv R, Bittel Barb, Browsky Dalton, Houghtaling Penny, Drummond Colin K, Desai Milind, Gillinov A Marc
1Department of Biomedical EngineeringCase Western Reserve UniversityClevelandOH44106USA.
2Heart and Vascular InstituteCleveland ClinicClevelandOH44106USA.
IEEE J Transl Eng Health Med. 2019 Dec 13;8:2700204. doi: 10.1109/JTEHM.2019.2950397. eCollection 2020.
Background Wearable wrist-monitors offer an unobtrusive way to acquire heart rate data in an efficient manner. Previous work in this field has focused on studying healthy subjects during exercise but has yet to assess the efficacy of these devices in patients suffering from common cardiac arrhythmias such as atrial fibrillation. Objective The objective of this pilot study was to assess the accuracy of the Apple Watch heart rate monitor in fifty patients experiencing atrial fibrillation compared to telemetry. Results Results from this pilot clinical study demonstrated a correlation coefficient of 0.7 between all readings on the Apple Watch and telemetry. Furthermore, the Apple Watch assessed heart rate more accurately in patients who were in atrial fibrillation than in those that were not (r = 0.86, patients in AF, vs. r = 0.64, patients not in AF). Clinical Impact The presented data from this pilot study suggests that caution should be noted before using the Apple Watch 4 wearable wrist monitor to monitor heart rate in patients with cardiac arrhythmias such as atrial fibrillation.
背景 可穿戴式腕部监测器提供了一种以高效方式获取心率数据的不引人注意的方法。该领域以前的工作主要集中在研究运动中的健康受试者,但尚未评估这些设备在患有常见心律失常(如心房颤动)的患者中的功效。目的 这项初步研究的目的是评估Apple Watch心率监测器在50名心房颤动患者中与遥测相比的准确性。结果 这项初步临床研究的结果表明,Apple Watch上的所有读数与遥测之间的相关系数为0.7。此外,Apple Watch在心房颤动患者中评估心率比非心房颤动患者更准确(r = 0.86,房颤患者,vs. r = 0.64,非房颤患者)。临床影响 这项初步研究提供的数据表明,在使用Apple Watch 4可穿戴式腕部监测器监测患有心房颤动等心律失常的患者的心率之前应谨慎。