Wei Yijuan, Jia Fei, Hou Ping, Zha Kaiji, Pu Shi, Gao Jianbo
Department of Radiology, the First Affiliated Hospital of Zhengzhou University, Zhengzhou, 450000, Henan, China.
Department of Radiation Oncology, the First Affiliated Hospital of Zhengzhou University, Zhengzhou, 450000, Henan, China.
Insights Imaging. 2020 Mar 6;11(1):32. doi: 10.1186/s13244-020-0836-1.
This study aimed to explore the performance of Revolution CT virtual monoenergetic images (VMI) combined with the multi-material artifact reduction (MMAR) technique in reducing metal artifacts in oral and maxillofacial imaging.
There were significant differences in image quality scores between VMI + MMAR images and VMI+MARS (multiple artifact reduction system) images at each monochromatic energy level (p = 0.000). Compared with the MARS technology, the MMAR technology further reduced metal artifacts and improved the image quality. At VMI and VMI, the SD, CNR, and AI in the Revolution CT group were significantly lower than in the Discovery CT, but no significant differences in these parameters were found between two groups at VMI, VMI, and VMI (p > 0.05). The attenuation was comparable between two groups at any energy level (p > 0.05).
Compared with the MARS reconstruction technique of Discovery CT, the MMAR technique of Revolution CT is better to reduce the artifacts of dental implants in oral and maxillofacial imaging, which improves the image quality and the diagnostic value of surrounding soft tissues.
本研究旨在探讨Revolution CT虚拟单能图像(VMI)联合多物质伪影减少(MMAR)技术在减少口腔颌面成像中金属伪影方面的性能。
在每个单色能量水平下,VMI + MMAR图像与VMI + MARS(多伪影减少系统)图像的图像质量评分存在显著差异(p = 0.000)。与MARS技术相比,MMAR技术进一步减少了金属伪影并提高了图像质量。在VMI和VMI时,Revolution CT组的标准差、对比噪声比和人工智能显著低于Discovery CT组,但在VMI、VMI和VMI时两组在这些参数上未发现显著差异(p > 0.05)。在任何能量水平下两组的衰减相当(p > 0.05)。
与Discovery CT的MARS重建技术相比,Revolution CT的MMAR技术在减少口腔颌面成像中牙种植体的伪影方面更好,提高了图像质量以及周围软组织的诊断价值。