Bafijari Denis, Benedetti Alberto, Stamatoski Aleksandar, Baftijari Florent, Susak Zoran, Veljanovski Darko
Dental Implant Center "Vita Dent", Tetovo, Republic of Macedonia.
University Clinic for Maxillofacial Surgery, Ss "Cyril and Methodius" University of Skopje, Skopje, Republic of Macedonia.
Open Access Maced J Med Sci. 2019 Dec 13;7(23):4110-4115. doi: 10.3889/oamjms.2019.716. eCollection 2019 Dec 15.
This study aimed to investigate and assess primary and secondary dental implant stability during the osseointegration period.
A total of 77 implants were placed in 42 patients with 26 males and 16 females. The study was conducted by comparing the resonance frequency analysis (RFA) values of the implants inserted in the lower jaw. RFA was done immediately after implant insertion and after 12 weeks. Results were statistically evaluated using SPSS Statistics for Windows, Version 7.1. Level of significance was set at P < 0.05.
Significant differences were detected between the primary and secondary stability values, respectively. Maximum RFA value of 88 and the minimum value of 52 were observed. Stability values increased during the following three months, and all implants were successfully integrated without complication.
Our results indicate and suggest that there is a strong linear correlation between implant stability and ISQ values that can be directly estimated by the RFA, especially in the posterior edentulous mandible. Osstell implant device could represent a useful tool which can be used to identify the risk for implant failure.
本研究旨在调查和评估骨结合期种植体的初期和二期稳定性。
共为42例患者植入77枚种植体,其中男性26例,女性16例。通过比较下颌植入种植体的共振频率分析(RFA)值开展本研究。在种植体植入后即刻及12周后进行RFA。使用Windows版SPSS Statistics 7.1对结果进行统计学评估。显著性水平设定为P < 0.05。
分别在初期和二期稳定性值之间检测到显著差异。观察到RFA最大值为88,最小值为52。在随后的三个月中稳定性值增加,且所有种植体均成功骨结合,无并发症发生。
我们的结果表明,种植体稳定性与可通过RFA直接估算的ISQ值之间存在很强的线性相关性,尤其是在下颌后牙区无牙颌。Osstell种植体装置可能是一种有用的工具,可用于识别种植失败风险。