Bazoukis George, Kollias George, Tse Gary, Saplaouras Athanasios, Mililis Panagiotis, Sakellaropoulou Antigoni, Kolokathis Angelos-Michail, Asvestas Dimitrios, Efremidis Michael, Sideris Antonios, Letsas Konstantinos P
Department of Cardiology, Electrophysiology Laboratory Evangelismos General Hospital of Athens Athens Greece.
Department of Medicine and Therapeutics Faculty of Medicine Chinese University of Hong Kong Hong Kong, SAR China.
Clin Case Rep. 2020 Jan 30;8(3):582-583. doi: 10.1002/ccr3.2682. eCollection 2020 Mar.
Reel syndrome occurs due to the rotation of the implantable device on its transverse axis with subsequent coiling of the leads around the pulse generator. Device interrogation and chest X-ray should be performed in any case of device malfunction.
卷轴综合征是由于植入式装置绕其横轴旋转,随后导线在脉冲发生器周围盘绕所致。在任何装置发生故障的情况下,都应进行装置问询和胸部X光检查。