Furukawa George T, Riddle John L, Bigge William R
Institute for Basic Standards, National Bureau of Standards, Washington, D.C. 20234.
J Res Natl Bur Stand A Phys Chem. 1973 May-Jun;77A(3):309-332. doi: 10.6028/jres.077A.019.
The reproducibility of the International Practical Temperature Scale of 1968 (IPTS-68) in the region of 13.81 K to 90.188 K using standard platinum resistance thermometers is discussed. The effect of errors above 0 °C on the calibration between 13 and 90 K is analyzed. Calibrations resulting from least squares and "fixed point" treatment of the data are compared. The results show that in general the NBS-IPTS-68 temperature scale maintained at the National Bureau of Standards (NBS) by reference standard platinum resistance thermometers is transferred to thermometers calibrated at the NBS within a few tenths of a mK between about 20 and 90.188 K and within about ±1 mK between 13.81 and 20 K. Agreements within a few mK are shown between the observed values of temperatures, employing platinum resistance thermometers calibrated in terms of the NBS-IPTS-68 scale, and the values of temperatures of the fixed points in the region 13.81 to 90.188 K given in the text of the IPTS-68.
讨论了使用标准铂电阻温度计在13.81 K至90.188 K范围内1968年国际实用温标(IPTS - 68)的复现性。分析了0°C以上的误差对13至90 K之间校准的影响。比较了对数据进行最小二乘法和“固定点”处理得到的校准结果。结果表明,一般来说,由国家标准局(NBS)的参考标准铂电阻温度计维持的NBS - IPTS - 68温标,在约20至90.188 K之间能以几十分之一毫开尔文的精度传递到在NBS校准的温度计上,在13.81至20 K之间精度约为±1毫开尔文。使用根据NBS - IPTS - 68温标校准的铂电阻温度计测得的温度观测值,与IPTS - 68文本中给出的13.81至90.188 K范围内固定点的温度值之间,显示出在几毫开尔文范围内的一致性。