Mangum B W
Clin Chem. 1983 Jul;29(7):1380-4.
In an investigation of the melting and freezing behavior of succinonitrile, the triple-point temperature was determined to be 58.0805 degrees C, with an estimated uncertainty of +/- 0.0015 degrees C relative to the International Practical Temperature Scale of 1968 (IPTS-68). The triple-point temperature of this material is evaluated as a temperature-fixed point, and some clinical laboratory applications of this fixed point are proposed. In conjunction with the gallium and ice points, the availability of succinonitrile permits thermistor thermometers to be calibrated accurately and easily on the IPTS-68.
在一项关于丁二腈熔化和凝固行为的研究中,确定其三相点温度为58.0805摄氏度,相对于1968年国际实用温标(IPTS - 68),估计不确定度为±0.0015摄氏度。该材料的三相点温度被评估为一个温度固定点,并提出了这个固定点的一些临床实验室应用。与镓点和冰点一起,丁二腈的可用性使热敏电阻温度计能够在IPTS - 68上准确且容易地进行校准。