J Res Natl Bur Stand A Phys Chem. 1973 Sep-Oct;77A(5):671-677. doi: 10.6028/jres.077A.039.
A low frost-point humidity generator has been developed at NBS to provide a capability for calibration, testing and research at very low levels of water vapor content in such gases as atmospheric air, carbon dioxide and nitrogen. The generator produces frost points from -30 to -100 °C at ambient pressures from 500 to 200,000 pascals (0.005 to 2 atm.). This is equivalent to mixing ratios of 4 × 10 to 51 grams of water vapor per kilogram of dry air and to vapor pressures of 1.4 × 10 to 38 pascals. The generated test gas can be fed to a test chamber with independent temperature control between +25 and -100 °C. The uncertainty of the frost point in the test chamber is estimated not to exceed 0.05 deg C. Intercomparisons with a frost-point hygrometer as well as two gravimetric checks gave results which in all cases agreed to within 0.2 deg C in frost-point temperature.
美国国家标准局(NBS)已研制出一种低霜点湿度发生器,用于在诸如大气、二氧化碳和氮气等气体中水蒸气含量极低的情况下进行校准、测试和研究。该发生器在500至200,000帕斯卡(0.005至2个大气压)的环境压力下可产生-30至-100°C的霜点。这相当于每千克干空气中水蒸气的混合比为4×10至51克,蒸气压为1.4×10至38帕斯卡。所产生的测试气体可输送到一个温度可在+25至-100°C之间独立控制的测试腔室。测试腔内霜点的不确定度估计不超过0.05°C。与霜点湿度计以及两次重量法检查的比对结果表明,在所有情况下,霜点温度的一致性在0.2°C以内。