一种用于妇科间质近距离放射治疗的新型双瓣阴道窥器施源器设计的评估

Evaluation of a new bi-valve vaginal speculum applicator design for gynecologic interstitial brachytherapy.

作者信息

Kunogi Hiroaki, Yamaguchi Nanae, Sasai Keisuke

机构信息

Department of Radiation Oncology, Juntendo University, Tokyo, Japan.

出版信息

J Contemp Brachytherapy. 2020 Feb;12(1):27-34. doi: 10.5114/jcb.2020.92406. Epub 2020 Feb 28.

Abstract

PURPOSE

We designed a bi-valve vaginal speculum high-dose-rate (HDR) interstitial gynecologic brachytherapy applicator. This allows for both a direct view of the cervix and image-guided brachytherapy applicator placement. The purpose of this study was to assess the validity of the new applicator.

MATERIAL AND METHODS

The applicator was designed to have a 25-mm arc, which can be spread transversely to 35-mm wide, with 10 insertion holes; it was produced using a stereolithographic printer with biocompatible Dental SG resin. For resin radiodensity was measured in Hounsfield units (HU) using computed tomography (CT). Comparing the new applicator with a conventional intracavitary applicator, we evaluated the treated volume (including dimensions of 100% isodose volume at the central axis), V/D/D for a hypothetical cervix (a 2-cm-long and 4-cm-diameter cylinder), and dose points of organs at risk (OARs) (at 25 and 30 mm from the tandem). Based on dose-volume histogram (DVH) analysis of the cervix and dose points of OARs, the range of tolerance for the percent dose difference in the prescription dose was set at 5%.

RESULTS

The mean radiodensity of the Dental SG resin, which was magnetic resonance imaging compatible, was 118 HU. Dimensions of the 100% isodose volume measured at the central axis were 4.4 × 6.6 × 7.4 cm for the new applicator and 4.3 × 6.0 × 7.7 cm for the intracavitary applicator. The 100% prescription dose volumes were 110 cc and 113 cc for the new and conventional applicator, respectively. The percent difference in the hypothetical cervix V, D, and D between the new and intracavitary applicator were within 5%. The percent differences in dose points of OARs at 25 and 30 mm between the new and conventional applicators were within 5%.

CONCLUSIONS

Our speculum applicator can reproduce a conventional pear-shaped dose distribution. Our current clinical practice will use this applicator, which can improve the patient's treatment results.

摘要

目的

我们设计了一种双瓣阴道窥镜高剂量率(HDR)间质妇科近距离放射治疗施源器。这既能直接观察宫颈,又能实现图像引导下施源器的放置。本研究的目的是评估这种新型施源器的有效性。

材料与方法

该施源器设计为具有25毫米的弧长,可横向扩展至35毫米宽,有10个插入孔;它是使用具有生物相容性的牙科SG树脂通过立体光刻打印机制作而成。对于树脂,使用计算机断层扫描(CT)以亨氏单位(HU)测量其放射密度。将新型施源器与传统腔内施源器进行比较,我们评估了治疗体积(包括中心轴上100%等剂量体积的尺寸)、针对假设宫颈(一个2厘米长、4厘米直径的圆柱体)的V/D/D,以及危及器官(OARs)的剂量点(在距串联器25和30毫米处)。基于宫颈的剂量体积直方图(DVH)分析和OARs的剂量点,处方剂量中百分比剂量差异的耐受范围设定为5%。

结果

与磁共振成像兼容的牙科SG树脂的平均放射密度为118 HU。新型施源器在中心轴处测量的100%等剂量体积尺寸为4.4×6.6×7.4厘米,腔内施源器为4.3×6.0×7.7厘米。新型和传统施源器的100%处方剂量体积分别为110立方厘米和113立方厘米。新型施源器与腔内施源器在假设宫颈的V、D和D方面的百分比差异在5%以内。新型和传统施源器在25和30毫米处OARs剂量点的百分比差异在5%以内。

结论

我们的窥镜施源器能够重现传统的梨形剂量分布。我们目前的临床实践将使用这种施源器,它可以改善患者的治疗效果。