Barlow N L, Auslander M C, Rines D, Stelmachowicz P G
Boys Town National Institute for Communication Disorders in Children, Omaha, Nebraska.
Ear Hear. 1988 Oct;9(5):243-7. doi: 10.1097/00003446-198810000-00003.
This study was designed to investigate the reliability of real-ear measurements of sound pressure level (SPL) and to compare these values with two coupler measures of SPL. A commercially available probe tube microphone system was used to measure real ear SPL in both children and adults. Test-retest reliability decreased as a function of frequency for both groups and, in general, was slightly poorer for the children. For both groups, coupler to real ear differences were larger for the 2 cm3 coupler than for the reduced volume coupler; however, no significant differences were observed between groups. In addition, a measure of ear canal volume was not found to be a good predictor of coupler to real ear discrepancies.