Clin Lab. 2020 Apr 1;66(4). doi: 10.7754/Clin.Lab.2019.190803.
We tested the uniformity and stability of candidate reference materials (RMs) for serum potassium and aimed to design RMs with better quality that can meet all clinical test requirements and effectively solve the quantity traceability transfer problem.
Three levels of frozen mixed serum potassium candidate RMs were prepared and packed in freezing tubes. RMs were determined in triplicate in 10 vials randomly selected from each level. A one-way analysis of variance was used to evaluate the uniformity using a ratio of the mean squares among groups to mean squares within groups F < F0.05 as the criteria. Stability was studied by synchronization; the short-term stability of the serum potassium in the transport conditions was observed for 30, 15, and 7 days at refrigeration (2 - 8°C), room temperature (18 - 25°C), and 37°C, respectively. By linear regression analysis of variance, the straight line was used as an empirical model. The criterion for judging is |b1| < t0.95, n - 2·sb1.
Based on the statistical analysis using SPSS 17, the F values for the homogeneity tests of each level of the frozen mixed serum potassium RMs were 0.247, 0.117, and 0.162. These values were less than F0.05 (9, 20) = 2.39. When the short-term stability of the serum potassium was observed for 30, 12, and 4 days at 2 - 8°C, room temperature, and 37°C, respectively, |b1| < t0.95, n - 2·sb1, and instability was not observed.
The three levels of the frozen mixed serum potassium RMs have good uniformity. At 2 - 8°C, room temperature, and 37°C, the stability can be ensured for at least 30, 12, and 4 days, respectively. The serum potassium candidate RMs are sufficiently stable under these transportation conditions.
我们测试了候选参考物质(RM)血清钾的均匀性和稳定性,旨在设计质量更好的 RM,以满足所有临床测试要求,并有效解决数量可追溯性传递问题。
制备了 3 个水平的冰冻混合血清钾候选 RM,并将其包装在冷冻管中。从每个水平中随机选择 10 个小瓶,对 RM 进行 3 次重复测定。使用组间均方与组内均方的比值 F < F0.05 作为标准,采用单向方差分析评估均匀性。通过同步研究稳定性,观察血清钾在冷藏(2-8°C)、室温(18-25°C)和 37°C 下运输条件下 30、15 和 7 天的短期稳定性。通过方差分析的线性回归,使用直线作为经验模型。判断标准为 |b1| < t0.95, n - 2·sb1。
基于 SPSS 17 的统计分析,冰冻混合血清钾 RM 各水平的 F 值分别为 0.247、0.117 和 0.162。这些值均小于 F0.05(9,20)= 2.39。当在 2-8°C、室温、37°C 下分别观察血清钾 30、12 和 4 天的短期稳定性时,|b1| < t0.95, n - 2·sb1,且未观察到不稳定性。
冰冻混合血清钾 RM 的 3 个水平具有良好的均匀性。在 2-8°C、室温、37°C 下,稳定性分别至少可保证 30、12 和 4 天。在这些运输条件下,血清钾候选 RM 足够稳定。