高分辨率阻抗测压法测得的食管基线阻抗与pH-阻抗监测测得的夜间平均基线阻抗相关。

Esophageal Baseline Impedance From High-resolution Impedance Manometry Correlates With Mean Nocturnal Baseline Impedance From pH-impedance Monitoring.

作者信息

Horton Anthony, Sullivan Brian, Charles Katie, McIntosh Thasha, Davis Andrea, Gellad Ziad, Shimpi Rahul, Gyawali C Prakash, Patel Amit

机构信息

Division of Gastroenterology, Duke University School of Medicine, Durham, NC, USA.

Durham Veterans Affairs Medical Center, Durham, NC, USA.

出版信息

J Neurogastroenterol Motil. 2020 Sep 30;26(4):455-462. doi: 10.5056/jnm19142.

Abstract

BACKGROUND/AIMS: Esophageal baseline impedance (BI) can be extracted from pH-impedance tracings as mean nocturnal baseline impedance (MNBI), and from high-resolution impedance manometry (HRIM), but it is unknown if values are similar between acquisition methods across HRIM manufacturers. We aim to assess correlations between MNBI and BI from HRIM (BI-HRIM) from 2 HRIM manufacturers in the setting of physiologic acid exposure time (AET).

METHODS

HRIM and pH-impedance monitoring demonstrating physiologic AET (< 4%) off proton pump inhibitors were required. BI-HRIM was extracted as the average from 5 cm and 10 cm above the lower esophageal sphincter. Distal BI-HRIM (DBI-HRIM) was also extracted from the most distal channel (Medtronic studies). MNBI was extracted from 6 channels. Concordance between BI-HRIM across manufacturers with MNBI was analyzed.

RESULTS

Thirty-six patients met the inclusion criteria (59.6 ± 1.7 years; 22% female; body mass index 30.5 ± 0.7; AET 1.6 ± 0.2%). Although MNBI was similar at all channels ( ≥ 0.18), Diversatek BI-HRIM was lower than Medtronic BI-HRIM ( = 0.003). Overall, BI-HRIM correlated with MNBI at corresponding recording sites, 7 cm and 9 cm ( < 0.05), but not at other sites ( ≥ 0.19). Pearson's correlations > 0.5 were seen at MNBI at 7 cm for both systems, and at 9 cm for Medtronic. DBI-HRIM correlated with MNBI at 3 cm and 5 cm ( < 0.03), but not at other locations ( > 0.1).

CONCLUSIONS

While numeric differences exist between manufacturers, BI-HRIM correlates with MNBI from corresponding channels in patients with physiologic AET. Comparison with AET elevation is needed to determine correlations between pathologic MNBI with BI-HRIM across manufacturers. The optimal HRIM channels from which BI values should be extracted also warrants further study.

摘要

背景/目的:食管基线阻抗(BI)可从pH-阻抗描记图中提取为平均夜间基线阻抗(MNBI),也可从高分辨率阻抗测压法(HRIM)中提取,但不同HRIM制造商的采集方法所得到的值是否相似尚不清楚。我们旨在评估在生理性酸暴露时间(AET)情况下,来自2家HRIM制造商的MNBI与HRIM的BI(BI-HRIM)之间的相关性。

方法

需要进行HRIM和pH-阻抗监测,以证明在停用质子泵抑制剂的情况下生理性AET(<4%)。BI-HRIM提取为食管下括约肌上方5 cm和10 cm处的平均值。还从最远端通道(美敦力研究)提取远端BI-HRIM(DBI-HRIM)。MNBI从6个通道提取。分析了不同制造商的BI-HRIM与MNBI之间的一致性。

结果

36例患者符合纳入标准(年龄59.6±1.7岁;女性占22%;体重指数30.5±0.7;AET 1.6±0.2%)。尽管所有通道的MNBI相似(≥0.18),但Diversatek BI-HRIM低于美敦力BI-HRIM(P = 0.0)。总体而言,BI-HRIM与相应记录部位(7 cm和9 cm)的MNBI相关(P<0.05),但在其他部位不相关(P≥0.19)。两个系统在7 cm处的MNBI以及美敦力在9 cm处的MNBI的皮尔逊相关性>0.5。DBI-HRIM与3 cm和5 cm处的MNBI相关(P<0.03),但在其他位置不相关(P>0.1)。

结论

虽然不同制造商之间存在数值差异,但在生理性AET患者中,BI-HRIM与相应通道的MNBI相关。需要与AET升高情况进行比较,以确定不同制造商之间病理性MNBI与BI-HRIM之间的相关性。提取BI值的最佳HRIM通道也值得进一步研究。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8ee6/7547185/812920f943e1/JNM-26-455-f1.jpg