Comparative biomechanical study of five systems for fixation of the coracoid transfer during the Latarjet procedure for treatment of anterior recurrent shoulder instability.

机构信息

Department of Orthopaedics and Traumatology, Sainte-Marguerite Hospital, Assistance Publique-Hôpitaux de Marseille, Marseille, France.

CNRS, ISM, Aix Marseille University, Marseille, France.

出版信息

Int Orthop. 2020 Sep;44(9):1767-1772. doi: 10.1007/s00264-020-04565-2. Epub 2020 May 12.

Abstract

PURPOSE

This work compares the biomechanical resistance of five modes of fixation coracoid bone-block fixation during Latarjet open-air or arthroscopic procedures. The hypothesis is that these systems are equivalent.

METHODS

Latarjet procedures were performed on cadavers, then the samples were subjected to an increasing tension until the fixation failed. Five systems were tested: two malleolar screws, one screw with washer, two 3.5-mm self-compressive screws, one 4-mm self-compressive screw associated with one 3-mm self-compressive screw, and endobutton. The main judgment criterion was the strength necessary for the failure of the fixation. The secondary criterion was the stiffness of the assembly.

RESULTS

The single malleolar screw fixing has a lower breaking threshold than other fixings. There is no difference in strength concerning the other systems. The average strength is greater than the stresses of a shoulder during daily life activities. There is no difference regarding the secondary criterion.

CONCLUSION

The use of a single screw is insufficient, but the other systems seems reliable. The use of small diameter self-compressive cannulated screws can provide a better result. This biomechanical work must be validated in clinical studies.

摘要

目的

本研究比较了 Latarjet 切开术或关节镜下 5 种喙突骨块固定方式的生物力学阻力。假设这些系统是等效的。

方法

对尸体进行 Latarjet 手术,然后对标本施加逐渐增加的张力,直到固定失败。测试了 5 种系统:两个骨螺钉、一个带垫圈的螺钉、两个 3.5mm 自攻加压螺钉、一个 4mm 自攻加压螺钉与一个 3mm 自攻加压螺钉、以及 Endobutton。主要判断标准是固定失败所需的强度。次要标准是组件的刚度。

结果

单骨螺钉固定的断裂阈值低于其他固定方式。其他系统的强度没有差异。平均强度大于日常生活中肩部的应力。次要标准没有差异。

结论

使用单螺钉是不够的,但其他系统似乎可靠。使用小直径自攻加压空心螺钉可以提供更好的结果。这项生物力学研究必须在临床研究中得到验证。