Huang Jian-Feng, Li Yan, Wang Ji-Guang
Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
Blood Press Monit. 2020 Aug;25(4):231-235. doi: 10.1097/MBP.0000000000000452.
To evaluate the accuracy of the Rossmax AC1000f oscillometric upper-arm blood pressure monitor in adults according to the British Hypertension Society Protocol.
Eighty-five subjects (male 42 and female 43) with a mean age of 48.1 years were recruited to fulfill the blood pressure distribution criteria of the British Hypertension Society Protocol in general population using the same arm sequential blood pressure measurement method. For each subject, we sequentially measured blood pressure using a mercury sphygmomanometer (two observers) and the AC1000f device (one supervisor). Data analysis was carried out according to the British Hypertension Society Protocol.
A total of 255 comparison pairs were obtained for analysis. The device achieved grade A. The percentage of blood pressure differences within 5, 10 and 15 mmHg was 66.7, 92.2 and 98.0%, respectively, for SBP, and 64.7, 90.2 and 98.4%, respectively, for DBP. The average (±SD) of the device-observer differences was -0.5 ± 6.0 mmHg for SBP and -1.4 ± 5.9 mmHg for DBP, respectively.
The automated upper-arm blood pressure monitor Rossmax AC1000f has passed the requirements of the British Hypertension Society Protocol, and hence can be recommended for blood pressure measurement in adults.
根据英国高血压学会协议评估Rossmax AC1000f上臂式电子血压计在成人中的准确性。
招募了85名受试者(男性42名,女性43名),平均年龄48.1岁,采用同一手臂连续血压测量方法,以满足英国高血压学会协议在普通人群中的血压分布标准。对于每名受试者,我们使用汞柱式血压计(两名观察者)和AC1000f设备(一名监督者)依次测量血压。根据英国高血压学会协议进行数据分析。
共获得255对比较数据用于分析。该设备达到A级。收缩压在5、10和15 mmHg范围内的血压差异百分比分别为66.7%、92.2%和98.0%,舒张压分别为64.7%、90.2%和98.4%。设备与观察者之间收缩压差异的平均值(±标准差)为-0.5±6.0 mmHg,舒张压为-1.4±5.9 mmHg。
Rossmax AC1000f自动上臂式血压计已通过英国高血压学会协议的要求,因此可推荐用于成人血压测量。