Aspects of instrumentation design for impedance imaging.

作者信息

Murphy D, Rolfe P

机构信息

Biomedical Engineering Centre, University of Oxford, UK.

出版信息

Clin Phys Physiol Meas. 1988;9 Suppl A:5-14. doi: 10.1088/0143-0815/9/4a/002.

Abstract

This paper takes the circuit for a general purpose impedance imaging data measurement system and analyses the importance of several components. For high accuracy it is shown that stray capacitances in the input circuitry are the major problem although the use of a phase sensitive demodulator can minimise this problem. It is shown that harmonic distortion in the constant current generator can cause errors which are, however, reduced if a linear multiplying demodulator is used rather than a switching demodulator.

摘要

本文以通用阻抗成像数据测量系统的电路为例,分析了几个组件的重要性。结果表明,对于高精度而言,尽管使用相敏解调器可以使该问题最小化,但输入电路中的杂散电容是主要问题。结果表明,恒流发生器中的谐波失真会导致误差,不过,如果使用线性乘法解调器而非开关解调器,误差会减小。