AJR Am J Roentgenol. 2020 Jul;215(1):159-164. doi: 10.2214/AJR.19.22492. Epub 2020 May 14.
Breast tissue expanders with magnetic ports are MRI unsafe, preventing patients from benefiting from the diagnostic capabilities of MRI. A tissue expander was recently developed with a radiofrequency identification (RFID) port used for needle location and expansion that may be acceptable for a patient undergoing MRI. The purpose of this investigation was to evaluate MRI issues using standardized techniques and well-accepted methods for this tissue expander with RFID port. The breast tissue expander with RFID port (Motiva Flora Tissue Expander, Establishment Labs) was assessed for magnetic field interactions (translational attraction and torque, 3 T), MRI-related heating (1.5 T/64 MHz and 3 T/128 MHz), artifacts (3 T), and functional changes associated with different MRI conditions (1.5 T/64 MHz and 3 T/128 MHz). Magnetic field interactions were minor (deflection angle of 2° and no torque) and thus will not pose a risk. At 1.5 T/64 MHz and 3 T/128 MHz, the highest temperature elevations (1.7°C and 1.9°C, respectively) were physiologically inconsequential. The tissue expander with RFID port exhibited relatively small artifacts on MRI. Exposures of the tissue expander with RFID port to different MRI conditions did not impact the ability to localize the RFID port or to read the electronic serial number. The findings indicated that this tissue expander with RFID port is "MR Conditional" for a patient referred for MRI at 1.5 T or 3 T. Importantly, the relatively small artifact associated with this implant offers potential advantages for patients undergoing MRI compared with tissue expanders that have magnetic ports that create substantial signal losses and distortions on MR images.
带有磁体端口的乳房组织扩张器对 MRI 是不安全的,使患者无法从 MRI 的诊断能力中获益。最近开发了一种带有射频识别 (RFID) 端口的组织扩张器,用于针的定位和扩张,这可能对接受 MRI 的患者是可以接受的。本研究的目的是使用标准化技术和为此带有 RFID 端口的组织扩张器接受的公认方法来评估 MRI 问题。带有 RFID 端口的乳房组织扩张器(Motiva Flora 组织扩张器,Establishment Labs)评估了磁场相互作用(平移吸引力和扭矩,3 T)、与 MRI 相关的加热(1.5 T/64 MHz 和 3 T/128 MHz)、伪影(3 T)以及与不同 MRI 条件相关的功能变化(1.5 T/64 MHz 和 3 T/128 MHz)。磁场相互作用很小(偏差角为 2°,没有扭矩),因此不会造成风险。在 1.5 T/64 MHz 和 3 T/128 MHz 下,最高温度升高(分别为 1.7°C 和 1.9°C)在生理上无足轻重。带有 RFID 端口的组织扩张器在 MRI 上显示出相对较小的伪影。带有 RFID 端口的组织扩张器暴露于不同的 MRI 条件下,不会影响定位 RFID 端口或读取电子序列号的能力。研究结果表明,这种带有 RFID 端口的组织扩张器在 1.5 T 或 3 T 进行 MRI 检查的患者中是“MR 条件”的。重要的是,与带有磁体端口的组织扩张器相比,该植入物产生的相对较小的伪影为接受 MRI 的患者提供了潜在的优势,因为带有磁体端口的组织扩张器会在 MR 图像上产生大量的信号丢失和失真。