Munhoz Alexandre Mendonça, Chala Luciano, Melo Giselle de, Azevedo Marques Neto Ary de, Tucunduva Tatiana
Breast Surgery Group, Plastic Surgery Division, Rua Mato Grosso, 306 cj.1706 Higienópolis ZIP, São Paulo, SP, 01239-040, Brazil.
Plastic Surgery Department - Hospital Moriah, Hospital Sírio-Libanês, Rua Mato Grosso, 306 cj.1706 Higienópolis ZIP, São Paulo, SP, 01239-040, Brazil.
Aesthetic Plast Surg. 2021 Dec;45(6):2645-2655. doi: 10.1007/s00266-021-02355-8. Epub 2021 Jun 1.
SmoothSilk implants (SSI) are the first generation of implants to incorporate a radio-frequency identification device (RFID-M), a non-invasive traceability system. Although the RFID-M is considered compatible with magnetic resonance imaging (MRI), the size of the artifact and its influence on breast tissue vary. This prospective study assessed safety and MRI issues in a cohort of breast reconstruction patients.
Forty-four SSI were used for breast reconstruction in patients undergoing treatment for breast cancer. All patients were evaluated for magnetic field interactions, MRI-related heating and artifacts in a 1.5-T MRI unit using standard T1/T2-weighted sequences utilized in clinical assessment of breast tissue/implants.
Mean patient age was 41.5 years (27-53ys) and body mass index was 28+-6.44 kg/m. In 18/22 patients (81.8%), mastectomies were unilateral. No patients reported local heat/discomfort. All implants showed RFID-M-related artifacts with an estimated mean volume in T1 of 42.9cm (26.2-63.6cm; SD±8.6 and 95% CI, 40.37-45.45) and in T2 of 60.5cm (35.4-97.2cm; SD±14.7 and 95% CI, 56.29-65.01). Artifact volume was smaller in T1 than in T2, to a statistically significant degree (p <0.001). There were no statistically significant differences in artifact volume according to surgical indication, breast side or implant volume. There were 4/44 (9%) cases of minor rotation (<45°). In all cases, adequate analysis of the breast tissue was performed.
The results demonstrate that SSI with RFID-M technology presented an artifact volume of 42.9cm and 60.5cm in T1 and T2 images, respectively. Our findings provide detailed information on the quality and location of MRI artifacts in a reconstructed cohort which can help guide clinical decision-making for patients. To our knowledge, this is the first time RFID-M breast implants have been prospectively evaluated for clinical and MRI issues in a cohort of reconstructive patients.
This journal requires that authors assign a level of evidence to each article. For a full description of these Evidence-Based Medicine ratings, please refer to the Table of Contents or the online Instructions to Authors www.springer.com/00266 .
SmoothSilk植入物(SSI)是第一代集成了射频识别装置(RFID-M)的植入物,这是一种非侵入性可追溯系统。尽管RFID-M被认为与磁共振成像(MRI)兼容,但伪影的大小及其对乳腺组织的影响各不相同。这项前瞻性研究评估了一组乳房重建患者的安全性和MRI相关问题。
44个SSI被用于乳腺癌治疗患者的乳房重建。所有患者均在1.5-T MRI设备中,使用临床评估乳腺组织/植入物时采用的标准T1/T2加权序列,评估磁场相互作用、MRI相关发热和伪影。
患者平均年龄为41.5岁(27 - 53岁),体重指数为28±6.44kg/m²。在22例患者中的18例(81.8%),乳房切除术为单侧。没有患者报告局部发热/不适。所有植入物均显示出与RFID-M相关的伪影,T1加权像中估计平均体积为42.9cm³(26.2 - 63.6cm³;标准差±8.6,95%置信区间,40.37 - 45.45),T2加权像中为60.5cm³(35.4 - 97.2cm³;标准差±14.7,95%置信区间,56.29 - 65.01)。T1加权像中的伪影体积比T2加权像中小,差异具有统计学意义(p < 0.001)。根据手术指征、乳房侧别或植入物体积,伪影体积无统计学显著差异。44例中有4例(9%)发生轻微旋转(<45°)。所有病例均对乳腺组织进行了充分分析。
结果表明,采用RFID-M技术的SSI在T1和T2图像中的伪影体积分别为42.9cm³和60.5cm³。我们的研究结果提供了关于重建队列中MRI伪影质量和位置的详细信息,有助于指导患者的临床决策。据我们所知,这是首次对一组重建患者中的RFID-M乳房植入物进行临床和MRI相关问题的前瞻性评估。
证据水平IV:本刊要求作者为每篇文章指定证据水平。有关这些循证医学评级的完整描述,请参阅目录或在线作者指南www.springer.com/00266 。