Smart textiles biotechnology for electrocardiogram monitoring in horses during exercise on treadmill: Validation tests.

机构信息

Department of Veterinary Sciences, University of Pisa, Pisa, Italy.

Department of Information Engineering, School of Engineering, University of Pisa, Pisa, Italy.

出版信息

Equine Vet J. 2021 Mar;53(2):373-378. doi: 10.1111/evj.13296. Epub 2020 Jul 7.

Abstract

BACKGROUND

There are several bioengineering solutions aimed at improving human health and welfare. Smart electrodes based on textile substrates have met the growing demand for comfort, reliability, and robustness when acquiring physiological signals.

OBJECTIVES

Given the importance of good quality electrocardiograms (ECG) in equine sports medicine, this study focuses on the validation of smart textile electrodes to acquire ECG signals in horses during treadmill exercise.

STUDY DESIGN

The performance of the smart textile electrodes is compared with standard silver/silver chloride (Ag/AgCl) electrodes in terms of signal quality.

METHODS

Five healthy Standardbred mares were fitted with two identical electronic systems for the simultaneous recording of ECGs during a standardised exercise test (SET) on a treadmill. One system was equipped with smart textile electrodes, whereas the second was equipped with standard Ag/AgCl electrodes. The Ag/AgCl electrodes were positioned on shaved skin with self-adhesive pads, and without (SET1) or with glue (SET2). The textile electrodes were positioned without shaving the skin. The Kurtosis (k) value for each ECG trace recorded was calculated as an index of ECG signal quality.

RESULTS

For the textile electrodes, k values were higher, and closer to ideal compared to Ag/AgCl electrodes. The median values of the Signal Quality Indexes (kSQI) were higher for textile compared to Ag/AgCl electrodes. These differences were significant in SET 2 (P < .001), but not in SET 1 (P = .08).

MAIN LIMITATIONS

This study was limited to treadmill exercise that did not include a rider or harness.

CONCLUSIONS

During treadmill exercise, textile electrodes are a practical solution for collecting good quality ECG traces.

摘要

背景

有几种生物工程解决方案旨在改善人类健康和福利。基于纺织基底的智能电极满足了人们对获取生理信号时舒适性、可靠性和坚固性日益增长的需求。

目的

鉴于高质量心电图(ECG)在赛马运动医学中的重要性,本研究侧重于验证智能纺织电极在马进行跑步机运动时获取 ECG 信号的能力。

研究设计

从信号质量的角度比较了智能纺织电极和标准的银/氯化银(Ag/AgCl)电极的性能。

方法

对 5 匹健康的标准种母马进行了测试,在跑步机上进行标准运动测试(SET)期间,使用两个相同的电子系统同时记录 ECG。一个系统配备了智能纺织电极,而另一个系统配备了标准的 Ag/AgCl 电极。Ag/AgCl 电极用自粘垫贴在剃过毛的皮肤上,而不使用(SET1)或使用胶水(SET2)。纺织电极在不剃毛的皮肤上定位。记录的每个 ECG 迹线的峰度(k)值被计算为 ECG 信号质量的指标。

结果

对于纺织电极,k 值更高,并且比 Ag/AgCl 电极更接近理想值。与 Ag/AgCl 电极相比,纺织品的信号质量指数(kSQI)的中位数更高。在 SET2 中(P<.001),这些差异具有统计学意义,但在 SET1 中(P=.08)则没有。

主要局限性

本研究仅限于不包括骑手或马具的跑步机运动。

结论

在跑步机运动中,纺织电极是一种实用的方法,可以采集到高质量的 ECG 迹线。