Objective Quantification of Image Quality and Optical Scatter Before and After Nd:YAG Capsulotomy Using a Double-Pass Technique.

作者信息

Zafar Abu-Bakar, Chu Richard C, Bishara Miranda N, Qazi Mujtaba A, Mahmoud Ashraf M, Pepose Jay S

机构信息

Department of Ophthalmology, Carle Foundation Hospital, Champaign, IL, USA.

Chu Eye Institute, Fort Worth, TX, USA.

出版信息

Clin Ophthalmol. 2020 May 26;14:1403-1411. doi: 10.2147/OPTH.S248286. eCollection 2020.

Abstract

PURPOSE

The purpose of this study is to evaluate and compare the correlation between changes in vision and HD Analyzer dual-pass metrics versus changes in vision and conventional subjective slit lamp gradings in pseudophakic patients with posterior capsular opacity undergoing neodymium:yttrium-aluminum-garnet (Nd:YAG) capsulotomy.

PATIENTS AND METHODS

High contrast (HC) and low contrast (LC) best spectacle-corrected distance visual acuity (BCVA) and HD Analyzer evaluation were prospectively performed on patients with mild-to-moderate posterior capsular opacification (PCO) and monofocal and accommodating intraocular lens implants. Differences between pre- and post-operative measurements were calculated, along with the correlation of HD Analyzer metrics and slit lamp grading to changes in visual acuity.

RESULTS

Following Nd:YAG capsulotomy (n=29), there was statistically significant improvement in HC-BCVA and LC-BCVA, decrease in optical scatter, and corresponding improvement in Strehl ratio and HD Analyzer values at all contrast levels tested (p≤0.05). Pearson test showed a high correlation between the improvement in HC-BCVA (r coefficient = 0.78) and LC-BCVA (r coefficient = 0.71) to the improvement in Objective Scatter Index (OSI). There was a higher correlation of change in HC-BCVA to pre-op OSI  (r=0.61) than to the subjective PCO grading score (r = 0.19). There was also a higher correlation of change in LC-BCVA to pre-op OSI (r = 0.49) than to subjective grading (r = 0.16).

CONCLUSION

The HD Analyzer provides objective measurements of forward light scatter (ie, light directed towards the retina) that can assist with both PCO grading and prediction of improvement of visual quality after YAG laser capsulotomy with higher accuracy than conventional slit lamp assessment based upon backscatter (ie, light traveling to the observer) in patients tested with monofocal and accommodating intraocular lens implants.

摘要

目的

本研究旨在评估和比较在接受钕:钇铝石榴石(Nd:YAG)晶状体囊切开术的后囊膜混浊假晶状体患者中,视力变化与HD分析仪双程测量指标之间的相关性,以及视力变化与传统主观裂隙灯分级之间的相关性。

患者与方法

对患有轻度至中度后囊膜混浊(PCO)且植入单焦点和可调节人工晶状体的患者,前瞻性地进行高对比度(HC)和低对比度(LC)最佳矫正远视力(BCVA)以及HD分析仪评估。计算术前和术后测量值之间的差异,以及HD分析仪指标和裂隙灯分级与视力变化的相关性。

结果

在Nd:YAG晶状体囊切开术后(n = 29),在所有测试的对比度水平下,HC-BCVA和LC-BCVA均有统计学上的显著改善,光学散射减少,斯特列尔比率和HD分析仪值相应提高(p≤0.05)。皮尔逊检验显示,HC-BCVA(r系数 = 0.78)和LC-BCVA(r系数 = 0.71)的改善与客观散射指数(OSI)的改善之间存在高度相关性。HC-BCVA变化与术前OSI的相关性(r = 0.61)高于与主观PCO分级评分的相关性(r = 0.19)。LC-BCVA变化与术前OSI的相关性(r = 0.49)也高于与主观分级的相关性(r = 0.16)。

结论

HD分析仪提供了向前光散射(即射向视网膜的光)的客观测量,与基于向后散射(即射向观察者的光)的传统裂隙灯评估相比,在接受单焦点和可调节人工晶状体植入的患者中,它能以更高的准确性辅助PCO分级和预测YAG激光晶状体囊切开术后视觉质量的改善。