一种移动家庭盆底肌肉训练器(Elvie Trainer)的可靠性和有效性研究。

Reliability and validity of a mobile home pelvic floor muscle trainer: The Elvie Trainer.

机构信息

Department of Mechanical Engineering, University of Ottawa, Ottawa, Ontario, Canada.

School of Rehabilitation Sciences, University of Ottawa, Ottawa, Ontario, Canada.

出版信息

Neurourol Urodyn. 2020 Aug;39(6):1717-1731. doi: 10.1002/nau.24439. Epub 2020 Jun 18.

Abstract

AIMS

Reliability and validity of force measurement and task detection by the Elvie Trainer were evaluated against an intravaginal dynamometer (IVD) and ultrasound (US) imaging.

METHODS

Women were recruited from local physiotherapy clinics. At the first visit, pelvic floor muscle (PFM) strength and tone were assessed manually. Women performed two sets of three repetitions of rest, PFM maximal voluntary contraction (MVC), and maximal Valsalva maneuver (MVM) tasks in supine and standing, with the Elvie Trainer in situ. Women performed another set of rest and MVC repetitions with a custom IVD in situ. At the second visit, PFM strength and tone were reassessed manually. Women performed two sets of three repetitions of the rest, PFM MVC, and MVM tasks in supine and standing, with the Elvie Trainer in situ. Concurrent US imaging was then acquired during a final set of PFM MVC and MVM repetitions in supine and standing, while the Elvie Trainer remained in situ. Reliability was evaluated using intraclass correlation coefficients. Validity was evaluated using Spearman's/Pearson's correlations and receiver operator characteristic curves.

RESULTS

Thirty women participated in the study. The Elvie Trainer MVC force outcomes exhibited excellent within-day and good between-day reliability, but were significantly lower than IVD measures, and exhibited poor relationships with IVD force outcomes. The Elvie Trainer was able to specify correct/incorrect performance of a PFM MVC.

CONCLUSIONS

The Elvie Trainer exhibits acceptable within-day and between-day reliability and can detect the correct performance of PFM MVCs; however, force measurements are not valid indicators of PFM strength and should not be used to measure outcomes.

摘要

目的

评估 Elvie Trainer 在测量力量和检测任务方面的可靠性和有效性,方法是将其与阴道内测力计(IVD)和超声(US)成像进行对比。

方法

从当地物理治疗诊所招募女性参与者。在第一次就诊时,通过手动评估盆底肌(PFM)的力量和张力。女性在仰卧和站立位置下,使用 Elvie Trainer 进行两组三次的休息、PFM 最大自主收缩(MVC)和最大瓦氏动作(MVM)任务,每组三次。女性在原位再次进行一组休息和 MVC 重复动作,使用定制的 IVD。在第二次就诊时,通过手动再次评估 PFM 的力量和张力。女性在仰卧和站立位置下,使用 Elvie Trainer 进行两组三次的休息、PFM MVC 和 MVM 任务,每组三次。然后,在仰卧和站立位置下的最后一组 PFM MVC 和 MVM 重复动作期间,同时进行连续 US 成像,而 Elvie Trainer 仍留在原位。使用组内相关系数评估可靠性。使用 Spearman's/Pearson's 相关性和受试者工作特征曲线评估有效性。

结果

共有 30 名女性参与了研究。Elvie Trainer 的 MVC 力量结果显示出良好的日内和良好的日间可靠性,但明显低于 IVD 测量值,并且与 IVD 力量结果相关性较差。Elvie Trainer 能够正确/不正确地指定 PFM MVC 的性能。

结论

Elvie Trainer 表现出可接受的日内和日间可靠性,并且能够检测 PFM MVC 的正确性能;然而,力量测量不是 PFM 强度的有效指标,不应用于测量结果。