Suppr超能文献

新型可穿戴设备测量肘部力量的可靠性和有效性。

Reliability and Validity of a Novel Wearable Device for Measuring Elbow Strength.

机构信息

Department of Mechanical Engineering, University of New Brunswick, Fredericton, NB E3B 5A3, Canada.

Institute of Biomedical Engineering, University of New Brunswick, Fredericton, NB E3B 5A3, Canada.

出版信息

Sensors (Basel). 2020 Jun 17;20(12):3412. doi: 10.3390/s20123412.

Abstract

Muscle strength is an important clinical outcome in rehabilitation and sport medicine, but options are limited to expensive but accurate isokinetic dynamometry (IKD) or inexpensive but less accurate hand-held dynamometers (HHD). A wearable, self-stabilizing, limb strength measurement device (LSMD) was developed to fill the current gap in portable strength measurement devices. The purpose of this study was to evaluate the reliability and validity of the LSMD in healthy adults. Twenty healthy adults were recruited to attend two strength testing sessions where elbow flexor and extensor strength was measured with the LSMD, with HHD and with IKD in random order, by two raters. Outcomes were intra-rater repeatability, inter-rater reproducibility and inter-session reproducibility using intra-class correlation coefficients (ICC). Limits of agreement and weighted least products regression were used to test the validity of the LSMD relative to the criterion standard (IKD), and calibration formulas derived to improve measurement fidelity. ICC values for the LSMD were >0.90 for all measures of reliability and for both muscle groups, but over-predicted extensor strength and under-predicted flexor strength. Validity was established by transforming the data with the criterion standard-based calibration. These data indicate that the LSMD is reliable and conditionally valid for quantifying strength of elbow flexors and extensors in a healthy adult population.

摘要

肌肉力量是康复和运动医学中的一个重要临床结果,但选择有限,昂贵但准确的等速测力(IKD)或便宜但不太准确的手持测力计(HHD)。为了填补便携式力量测量设备的当前空白,开发了一种可穿戴的、自稳定的肢体力量测量装置(LSMD)。本研究的目的是评估 LSMD 在健康成年人中的可靠性和有效性。招募了 20 名健康成年人参加两次力量测试会议,其中使用 LSMD、HHD 和 IKD 以随机顺序由两名评估者测量肘屈肌和伸肌的力量。使用组内相关系数(ICC)评估结果为内部评估者重复性、内部评估者重现性和内部测试间重现性。使用界限和加权最小二乘法回归来测试 LSMD 相对于标准(IKD)的有效性,并推导校准公式以提高测量准确性。对于所有可靠性测量和两个肌肉群,LSMD 的 ICC 值均>0.90,但过度预测了伸肌力量,低估了屈肌力量。通过使用基于标准的校准转换数据来确定有效性。这些数据表明,LSMD 可靠且有条件地有效,可用于量化健康成年人的肘屈肌和伸肌力量。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a125/7349842/f3d1e20fd139/sensors-20-03412-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验