Haemolysis induced by mechanical circulatory support devices: unsolved problems.

作者信息

Köhne Inge

机构信息

Department for Health Services Research, Carl von Ossietzky University of Oldenburg, Oldenburg, Germany.

出版信息

Perfusion. 2020 Sep;35(6):474-483. doi: 10.1177/0267659120931307. Epub 2020 Jun 23.

Abstract

Since the use of continuous flow blood pumps as ventricular assist devices is standard, the problems with haemolysis have increased. It is mainly induced by shear stress affecting the erythrocyte membrane. There are many investigations about haemolysis in laminar and turbulent blood flow. The results defined as threshold levels for the damage of erythrocytes depend on the exposure time of the shear stress, but they are very different, depending on the used experimental methods or the calculation strategy. Here, the results are resumed and shown in curves. Different models for the calculation of the strengths of erythrocytes are discussed. There are few results reported about tests of haemolysis in blood pumps, but some theoretical approaches for the design of continuous flow blood pumps according to low haemolysis have been investigated within the last years.

摘要

由于将连续流式血泵用作心室辅助装置已成为标准做法,溶血问题有所增加。这主要是由影响红细胞膜的剪切应力引起的。关于层流和湍流血液流动中的溶血有许多研究。定义为红细胞损伤阈值水平的结果取决于剪切应力的暴露时间,但根据所使用的实验方法或计算策略,它们差异很大。在此,将结果汇总并以曲线形式展示。讨论了计算红细胞强度的不同模型。关于血泵中溶血测试的报道结果很少,但在过去几年中已经研究了一些根据低溶血设计连续流式血泵的理论方法。