Iida H, Higano S, Tomura N, Shishido F, Kanno I, Miura S, Murakami M, Takahashi K, Sasaki H, Uemura K
Research Institute for Brain and Blood Vessels, Akita, Japan.
J Cereb Blood Flow Metab. 1988 Apr;8(2):285-8. doi: 10.1038/jcbfm.1988.60.
The tracer appearance time relative to the radial artery-sampling site has been evaluated in six brain locations in five human subjects using dynamic positron emission tomography (PET) following the bolus injection of H2(15)O. There was a maximum difference of +/- 2 s from the average in each location. To globally adjust the timing difference between the measured arterial curve and the PET scan, a correction method was developed based on a nonlinear least-squares fitting procedure. This new technique determined the global time delay with an accuracy of +/- 0.5 s. On the other hand, the linear backward extrapolation method resulted in a systematic error of 4 s.
在5名人类受试者中,通过静脉注射H2(15)O后采用动态正电子发射断层扫描(PET)技术,评估了示踪剂相对于桡动脉采样部位在六个脑区的出现时间。每个脑区与平均值的最大差异为±2秒。为了整体调整测量的动脉曲线与PET扫描之间的时间差异,基于非线性最小二乘拟合程序开发了一种校正方法。这项新技术确定的整体时间延迟精度为±0.5秒。另一方面,线性向后外推法导致了4秒的系统误差。