Hemorrhagic shock is the first cause of preventable death in combat. Evacuations of wounded by aircraft are increasingly used and severely injured patients can spend consequent time in the air, mostly during strategic evacuation. In these situations, monitoring of blood coagulation may be pivotal in the management of blood product transfusion. Viscoelastic-guided transfusion is relevant in these situations. However, evaluation of these devices used in aircraft is lacking, especially the impact of decreased atmospheric pressure. The aim of this study is to evaluate the performance of an easy-to-carry viscoelastic system (TEG 6s, Haemonetics).
First, TEG 6s repeatability, reproducibility, and correlation with chronometric methods and TEG-5000 were assessed on quality controls, healthy volunteers, and patients. Secondly, we tested the influence of vibrations and altitude on TEG 6s parameters (0ft vs. 8000 ft = 2428 m) and on quality control samples (normal and hypocoagulable).
TEG 6s exhibited good correlation with the reference method and TEG 5000. Repeatability and reproducibility CVs were satisfactory. The tests performed in the hypobaric chamber revealed that performance at 0 ft and 8000 ft (2428 m) for 9 out of 13 parameters was not significantly different. However, we showed a significant increasing of CRT.Alpha (p = 0.049), CK.Alpha, CK.MA (p < 0.001 and p < 0.01, respectively) and CFF.MA increased (p < 0.05).
Our study provides proof of concept to validate testing in an actual aeromedical situation. Indeed, TEG 6s appears to ease of use, resistance to high altitude conditions, and reliability on healthy humans. It is necessary to carry out a study on hemorrhagic injured patients in an aircraft.
失血性休克是战斗中可预防死亡的首要原因。越来越多地使用飞机来转移伤员,重伤员在转移过程中可能会在飞机上停留很长时间,尤其是在战略转移中。在这些情况下,监测凝血功能对于血液制品输注的管理可能至关重要。黏弹性指导的输血在这些情况下是相关的。然而,在飞机上使用这些设备的评估是缺乏的,特别是在大气压力降低的情况下。本研究的目的是评估一种易于携带的黏弹性系统(TEG 6s,Haemonetics)的性能。
首先,我们评估了 TEG 6s 在质量控制、健康志愿者和患者身上的重复性、再现性以及与计时法和 TEG-5000 的相关性。其次,我们测试了振动和海拔高度对 TEG 6s 参数(0 英尺与 8000 英尺=2428 米)和质量控制样本(正常和低凝)的影响。
TEG 6s 与参考方法和 TEG 5000 具有良好的相关性。重复性和再现性 CV 令人满意。在低压舱中进行的测试显示,在 0 英尺和 8000 英尺(2428 米)的 13 个参数中的 9 个参数的性能没有显著差异。然而,我们发现 CRT.Alpha(p=0.049)、CK.Alpha、CK.MA(p<0.001 和 p<0.01,分别)和 CFF.MA 显著增加(p<0.05)。
我们的研究提供了在实际航空医疗情况下进行测试的概念验证。事实上,TEG 6s 易于使用,能够抵抗高空条件,在健康人群中具有可靠性。有必要在飞机上对出血性损伤患者进行研究。