Sensors (Basel). 2020 Jun 27;20(13):3613. doi: 10.3390/s20133613.
Measurement of carbon dioxide (CO) in medical applications is a well-established method for monitoring patient's pulmonary function in a noninvasive way widely used in emergency, intensive care, and during anesthesia. Even in extracorporeal-life support applications, such as Extracorporeal Carbon Dioxide Removal (ECCOR), Extracorporeal Membrane Oxygenation (ECMO), and cardiopulmonary by-pass (CPB), measurement of the CO concentration in the membrane oxygenator exhaust gas is proven to be useful to evaluate the treatment progress as well as the performance of the membrane oxygenator. In this paper, we present a new optical sensor specifically designed for the measurement of CO concentration in oxygenator exhaust gas. Further, the developed sensor allows measurement of the gas flow applied to the membrane oxygenator as well as the estimation of the CO removal rate. A heating module is implemented within the sensor to avoid water vapor condensation. Effects of temperature on the sensor optical elements of the sensors are disclosed, as well as a method to avoid signal-temperature dependency. The newly developed sensor has been tested and compared against a reference device routinely used in clinical practice in both laboratory and in vivo conditions. Results show that sensor accuracy fulfills the requirements of the ISO standard, and that is suitable for clinical applications.
在医疗应用中,二氧化碳(CO)的测量是一种成熟的方法,可无创监测患者的肺功能,广泛应用于急诊、重症监护和麻醉期间。即使在体外生命支持应用中,如体外 CO2 去除(ECCOR)、体外膜氧合(ECMO)和心肺旁路(CPB),测量膜式氧合器废气中的 CO 浓度也被证明有助于评估治疗进展和膜式氧合器的性能。在本文中,我们提出了一种专门用于测量氧合器废气中 CO 浓度的新型光学传感器。此外,所开发的传感器可测量应用于膜式氧合器的气体流量以及 CO 去除率的估计。传感器内实现了一个加热模块,以避免水蒸气冷凝。本文还揭示了温度对传感器光学元件的影响,以及一种避免信号-温度相关性的方法。新开发的传感器已在实验室和体内条件下针对临床实践中常规使用的参考设备进行了测试和比较。结果表明,传感器的准确性满足 ISO 标准的要求,适用于临床应用。