[使用斜孔准直器进行头部单光子发射计算机断层扫描的技术研究]

[Technical study for SPECT of the head using slant hole collimator].

作者信息

Ikeda H, Hamada K, Kosakai K, Omura M, Shimonishi Y, Ochi H

机构信息

Department of Radiology, Osaka City University Hospital, Japan.

出版信息

Radioisotopes. 1988 Apr;37(4):221-4. doi: 10.3769/radioisotopes.37.4_221.

Abstract

The performance of a 160 keV slant hole collimator (160 keV SH) was compared with that of a 140 keV high resolution collimator (140 keV HR) for SPECT imaging of the head using 99mTc. The minimum radius of revolution was reduced from about 23 cm to 13 cm using 160 keV SH. Relative sensitivity of the 160 keV SH was 108% of that of the 140 keV HR. Compared with SPECT using 140 keV HR, 160 keV SH SPECT yielded approximately a 15% increase in resolution. The 160 keV SH SPECT images of phantom were superior in resolution and contrast to that of 140 keV HR SPECT. We concluded that high quality SPECT images of the head can be obtained from rotating gamma cameras equipped with slant hole collimators.

摘要

使用99mTc对头进行SPECT成像时,将160 keV斜孔准直器(160 keV SH)的性能与140 keV高分辨率准直器(140 keV HR)的性能进行了比较。使用160 keV SH时,最小旋转半径从约23 cm减小到13 cm。160 keV SH的相对灵敏度为140 keV HR的108%。与使用140 keV HR的SPECT相比,160 keV SH SPECT的分辨率提高了约15%。模型的160 keV SH SPECT图像在分辨率和对比度方面优于140 keV HR SPECT。我们得出结论,配备斜孔准直器的旋转γ相机可以获得高质量的头部SPECT图像。

相似文献

8
Performance of a multislice fan beam collimator for SPECT imaging of the head.
Ann Nucl Med. 1991 Sep;5(3):117-21. doi: 10.1007/BF03164624.
9
Evaluation of a novel collimator for molecular breast tomosynthesis.新型乳腺断层合成准直器的评估。
Med Phys. 2017 Nov;44(11):5740-5748. doi: 10.1002/mp.12564. Epub 2017 Oct 13.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验