Ikeda H, Hamada K, Kosakai K, Omura M, Shimonishi Y, Ochi H
Department of Radiology, Osaka City University Hospital, Japan.
Radioisotopes. 1988 Apr;37(4):221-4. doi: 10.3769/radioisotopes.37.4_221.
The performance of a 160 keV slant hole collimator (160 keV SH) was compared with that of a 140 keV high resolution collimator (140 keV HR) for SPECT imaging of the head using 99mTc. The minimum radius of revolution was reduced from about 23 cm to 13 cm using 160 keV SH. Relative sensitivity of the 160 keV SH was 108% of that of the 140 keV HR. Compared with SPECT using 140 keV HR, 160 keV SH SPECT yielded approximately a 15% increase in resolution. The 160 keV SH SPECT images of phantom were superior in resolution and contrast to that of 140 keV HR SPECT. We concluded that high quality SPECT images of the head can be obtained from rotating gamma cameras equipped with slant hole collimators.
使用99mTc对头进行SPECT成像时,将160 keV斜孔准直器(160 keV SH)的性能与140 keV高分辨率准直器(140 keV HR)的性能进行了比较。使用160 keV SH时,最小旋转半径从约23 cm减小到13 cm。160 keV SH的相对灵敏度为140 keV HR的108%。与使用140 keV HR的SPECT相比,160 keV SH SPECT的分辨率提高了约15%。模型的160 keV SH SPECT图像在分辨率和对比度方面优于140 keV HR SPECT。我们得出结论,配备斜孔准直器的旋转γ相机可以获得高质量的头部SPECT图像。