Res Q Exerc Sport. 2021 Sep;92(3):549-558. doi: 10.1080/02701367.2020.1759768. Epub 2020 Jul 7.
: The present study analyzed peripheral blood oxygen saturation (SO) and heart rate (HR) measurements taken on the Garmin fēnix® 5X Plus watch, comparing them to measurements taken on a standard medical-grade pulse oximeter during normobaric hypoxia exposure under resting conditions. : Thirteen women (mean ± SD: Age 20 ± 1 years, height 165 ± 5 cm, mass, 67 ± 9 kg) and ten men (mean ± SD: Age 21 ± 3 years, height 177 ± 6 cm, mass 78 ± 11 kg) sat inside a customized environmental chamber while the fraction of inspired oxygen (FO) was adjusted to simulate altitudes of 12,000; 10,000; 8,000; 6,000; and 900 ft. The novel commercial device (Garmin fēnix®) and a medical-grade pulse oximeter (Nonin® 7500) were used to measure SO and HR in triplicate at each simulated altitude. Bland-Altman analyses were used to assess differences between methods. : Bland-Altman analysis indicated 3.3% bias for SO measurements taken on the Garmin fēnix® at 12,000 ft of simulated altitude (limits of agreement: -1.9-8.6%). Mean differences in SO measurements were smaller at the remaining simulated altitudes, where bias measurements ranged from 0.7% to 0.8%. The Garmin fēnix® also underestimated heart rate, but those discrepancies were minimal (bias measurements at all simulated altitude exposures were < 1.0 bpm). : With the exception of readings taken at 12,000 ft of simulated altitude, the Garmin fēnix® exhibits minimal overestimation of SO and minimal underestimation of HR during simulated altitude exposure. These data suggest the Garmin fēnix® watch may be a viable method to monitor SO and HR under most ambient environmental conditions.
本研究分析了 Garmin fēnix® 5X Plus 手表在常压低氧环境下静息状态下测量的外周血氧饱和度 (SO) 和心率 (HR),并将其与标准医用脉搏血氧仪的测量结果进行了比较。 13 名女性(平均值±标准差:年龄 20±1 岁,身高 165±5cm,体重 67±9kg)和 10 名男性(平均值±标准差:年龄 21±3 岁,身高 177±6cm,体重 78±11kg)坐在定制的环境室内,同时调整吸入氧分数(FO)以模拟海拔 12000 英尺;10000 英尺;8000 英尺;6000 英尺和 900 英尺。新型商业设备(Garmin fēnix®)和医用脉搏血氧仪(Nonin® 7500)用于在每个模拟海拔高度重复测量三次 SO 和 HR。 Bland-Altman 分析用于评估两种方法之间的差异。 Bland-Altman 分析表明,在模拟海拔 12000 英尺处,Garmin fēnix® 的 SO 测量值存在 3.3%的偏差(一致性界限:-1.9-8.6%)。在其余模拟海拔高度下,SO 测量值的平均差异较小,其中偏差测量值在 0.7%至 0.8%之间。Garmin fēnix® 还低估了心率,但这些差异很小(在所有模拟海拔暴露中,偏差测量值均<1.0 bpm)。 除了在模拟海拔 12000 英尺处的读数外,Garmin fēnix® 在模拟海拔暴露期间对 SO 的测量值表现出最小的高估和对 HR 的最小低估。这些数据表明,Garmin fēnix® 手表在大多数环境条件下可能是监测 SO 和 HR 的可行方法。