Abstract

Lung diseases have high mortality and morbidity, with an important impact on quality of life. Hypoxemic patients are advised to use oxygen therapy to prolong their survival, but high oxygen saturation (SpO) levels can also have negative effects. Pulse oximeters are the most common way to assess oxygen levels and guide medical treatment. This study aims to assess whether wearable devices can provide precise SpO measurements when compared to commercial pulse oximeters. This is a cross-section study with 100 patients with chronic obstructive pulmonary disease and interstitial lung disease from an outpatient pneumology clinic. SpO and heart rate data were collected with an Apple Watch Series 6 (Apple) and compared to two commercial pulse oximeters. The Bland-Altman method and interclass correlation coefficient were used to compare their values. We observed strong positive correlations between the Apple Watch device and commercial oximeters when evaluating heart rate measurements (r = 0.995, p < 0.001) and oximetry measurements (r = 0.81, p < 0.001). There was no statistical difference in the evaluation of skin color, wrist circumference, presence of wrist hair, and enamel nail for SpO and heart rate measurements in Apple Watch or commercial oximeter devices (p > 0.05). Apple Watch 6 is a reliable way to obtain heart rate and SpO in patients with lung diseases in a controlled environment.

摘要

肺部疾病死亡率和发病率高,对生活质量有重要影响。低氧血症患者建议使用氧疗来延长生存时间,但高血氧饱和度 (SpO) 水平也可能产生负面影响。脉搏血氧仪是评估氧气水平和指导治疗的最常用方法。本研究旨在评估与商业脉搏血氧仪相比,可穿戴设备在测量 SpO 时是否能提供更精确的测量值。这是一项横断面研究,共纳入了 100 名来自门诊肺病诊所的慢性阻塞性肺疾病和间质性肺病患者。使用 Apple Watch Series 6(苹果公司)收集 SpO 和心率数据,并与两个商业脉搏血氧仪进行比较。采用 Bland-Altman 法和组内相关系数比较它们的值。我们观察到 Apple Watch 设备与商业血氧仪在评估心率测量(r=0.995,p<0.001)和血氧测量(r=0.81,p<0.001)时呈强正相关。在评估 Apple Watch 或商业血氧仪设备的皮肤颜色、手腕周长、手腕毛发存在情况和指甲珐琅质对 SpO 和心率测量值时,差异无统计学意义(p>0.05)。在控制环境下,Apple Watch 6 是一种可靠的获取肺病患者心率和 SpO 的方法。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/40e4/8460792/ffd9580e3aba/41598_2021_98453_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验