使用红外显微镜系统(TEST1)测量红细胞沉降率时,预分析储存的影响。

Impact of preanalytical storage on the measurement of erythrocyte sedimentation rate using an infrared microphotometer system (TEST1).

机构信息

LABOSUD laboratory (Inovie member), Montpellier, France.

出版信息

Scand J Clin Lab Invest. 2020 Oct;80(6):523-524. doi: 10.1080/00365513.2020.1786887. Epub 2020 Jul 9.

Abstract

This study examined the influence of temperature and time on the pre-analytical stability of the erythrocyte sedimentation rate (ESR) measured on a TEST1 system. The first experiment included 102 samples stored at room temperature and the second experiment included 112 subjects and investigated refrigerated (2-8 °C) storage. Our study showed a stable ESR results at room temperature (15-25 °C) up to 8 h ( = 0.512). Samples stored at 2-8 °C for 24 h were stable ( = 0.280) for 24 h.

摘要

本研究考察了温度和时间对 TEST1 系统测量的红细胞沉降率(ESR)的预分析稳定性的影响。第一个实验包括 102 个在室温下储存的样本,第二个实验包括 112 个样本并研究了冷藏(2-8°C)储存。我们的研究表明,室温(15-25°C)下的 ESR 结果在 8 小时内稳定( = 0.512)。2-8°C 储存 24 小时的样本在 24 小时内稳定( = 0.280)。