Zeng Xiangfei, Wang Xiaoxia, Chen Huifang, Liu Daihong, Shi Jinfang, Zhang Jing, Li Xiaoqin, Zhang Liyin, Yang Yi, Zhang Jiuquan
From the Department of Radiology, Chongqing University Cancer Hospital & Chongqing Cancer Hospital, & Chongqing Cancer Institute, Chongqing, P.R. China.
J Comput Assist Tomogr. 2020 Nov/Dec;44(6):921-927. doi: 10.1097/RCT.0000000000001063.
The objective of this study was to evaluate the image quality of monoenergetic images (MEIs (+)) acquired from dual-energy computed tomography with low-concentration and low-flow-rate contrast media for the arterial supply to the nipple-areola complex (NAC) in breast cancer compared with conventional computed tomography angiography (CTA).
We enrolled 25 patients (MEI (+)300 group, 300 mg/mL and 2.5 mL/s of contrast media) and 23 patients (CTA370 group, 370 mg/mL and 3.5 mL/s of contrast media) for assessing NAC blood supply angiography. The image quality of the 2 groups was evaluated objectively and subjectively.
The 40 keV MEI (+)300 demonstrated higher attenuation and contrast-to-noise ratio than CTA370 group (P < 0.001). The subjective image quality and visualization of the arteries were comparable between 2 groups.
The 40 keV MEI (+)300 acquired from dual-energy computed tomography can achieve comparable image quality of arterial supply to NAC with low-concentration and low-flow-rate contrast media in breast cancer compared with CTA370.
本研究的目的是评估与传统计算机断层扫描血管造影(CTA)相比,采用低浓度、低流速造影剂的双能计算机断层扫描获取的单能量图像(MEIs(+))对乳腺癌乳头乳晕复合体(NAC)动脉供血的图像质量。
我们纳入了25例患者(MEI(+)300组,造影剂浓度为300mg/mL,流速为2.5mL/s)和23例患者(CTA370组,造影剂浓度为370mg/mL,流速为3.5mL/s)来评估NAC供血血管造影。对两组的图像质量进行客观和主观评估。
40keV的MEI(+)300显示出比CTA370组更高的衰减和对比噪声比(P<0.001)。两组之间主观图像质量和动脉可视化程度相当。
与CTA370相比,双能计算机断层扫描获取的40keV的MEI(+)300在使用低浓度、低流速造影剂时,对乳腺癌NAC动脉供血能实现相当的图像质量。