To investigate the feasibility of low-concentration contrast media (LC-CM) in cerebral and cervical dual-energy CT angiography (DE-CTA) using an advanced monoenergetic (Mono+) reconstruction technique.
Sixty-five consecutive patients prospectively selected to undergo cerebral and cervical DE-CTA were randomised into two groups: 32 patients (63.7 ± 9.7 years) in the high-concentration contrast medium (HC-CM) group with iopromide 370 and 33 patients (60.7 ± 10.8 years) in the low-concentration contrast medium (LC-CM) group with iodixanol 270. Traditional monoenergetic (Mono) and Mono+ images from 40 to 100 keV levels (at 10-keV intervals) and the standard mixed (Mixed, 120 kVp equivalent) images were reconstructed. Subjective image quality parameters included the contrast-to-noise ratio (CNR) and objective image quality parameters were evaluated and compared between the two groups.
The 40-keV Mono+ images in the LC-CM group showed comparable objective CNR (common carotid arteries: 83.7 ± 24.5 vs. 78.1 ± 23.2; internal carotid arteries: 82.2 ± 26.8 vs. 76.8 ± 24.1; middle cerebral arteries: 72.5 ± 24.6 vs. 70.6 ± 19.2; all p > 0.05) and subjective image scores (3.95 ± 0.19 vs. 3.83 ± 0.35; p > 0.05) compared with Mixed images in the HC-CM group.
The Mono+ reconstruction technique could reduce the concentration of iodinated CM in the diagnosis of cerebral and cervical angiography.
• Mono+ shows decreased noise and superior CNR compared with Mono. • The 40-keV Mono+ images show the highest CNR in the LC-CM group. • The Mono+ reconstruction technique could reduce the concentration of iodinated CM.
利用先进的单能量(Mono+)重建技术,研究低浓度对比剂(LC-CM)在脑颈双能 CT 血管造影(DE-CTA)中的可行性。
前瞻性选择 65 例连续行脑颈 DE-CTA 的患者,随机分为两组:高浓度对比剂(HC-CM)组 32 例(63.7±9.7 岁),碘普罗胺 370;低浓度对比剂(LC-CM)组 33 例(60.7±10.8 岁),碘昔醇 270。分别对 40~100keV 水平(10keV 间隔)的传统单能量(Mono)和 Mono+图像以及标准混合(Mixed,120kVp 等效)图像进行重建。主观图像质量参数包括对比噪声比(CNR),客观图像质量参数在两组间进行评估和比较。
LC-CM 组 40keV Mono+图像的客观 CNR(颈总动脉:83.7±24.5 比 78.1±23.2;颈内动脉:82.2±26.8 比 76.8±24.1;大脑中动脉:72.5±24.6 比 70.6±19.2;均 p>0.05)和主观图像评分(3.95±0.19 比 3.83±0.35;p>0.05)与 HC-CM 组的 Mixed 图像相似。
Mono+重建技术可降低脑颈血管造影诊断中碘对比剂的浓度。
•Mono+与 Mono 相比,噪声降低,CNR 更高。•LC-CM 组 40keV Mono+图像显示出最高的 CNR。•Mono+重建技术可降低碘对比剂的浓度。