Abstract

PURPOSE

Dedicated post-processing of dual-energy computed tomography angiography (DE-CTA) datasets has been shown to allow for increased vascular contrast. The goal of our study was to define optimal window settings for displaying virtual monoenergetic images (VMI) reconstructed from dual-energy carotid and cerebrovascular DE-CTA.

METHODS

Fifty-seven patients who underwent clinically-indicated carotid and cerebrovascular third-generation dual-source DE-CTA were retrospectively evaluated. Standard linearly-blended (M_0.6), 70-keV traditional VMI (M70), and 40-keV noise-optimized VMI (M40+) reconstructions were analyzed. For M70 and M40+ datasets, the subjectively best window setting (width and level, B-W/L) was independently determined by two observers and subsequently related with aortic arch attenuation to calculate optimized values (O-W/L) using linear regression. Subjective evaluation of image quality (IQ) between W/L settings were assessed by two additional readers. Repeated measures analysis of variance were performed to compare W/L settings and IQ indices between M_0.6, M70, and M40 + .

RESULTS

B-W/L and O-W/L for M70 were 580/210 and 560/200, and for M40+ were 1630/570 and 1560/550, respectively, higher than standard DE-CTA W/L settings (450/100). Highest subjective scores were observed for M40+ regarding overall IQ (all p < 0.001).

CONCLUSION

Application of O-W/L settings is mandatory to optimize subjective IQ of VMI reconstructions of DE-CTA.

摘要

目的

已经证实,双能计算机断层血管造影术(DE-CTA)数据集的专用后处理可以提高血管对比度。我们的研究目的是为了确定显示从双能颈动脉和脑血管 DE-CTA 重建的虚拟单能量图像(VMI)的最佳窗口设置。

方法

回顾性评估了 57 例接受临床指征明确的颈动脉和脑血管第三代双源 DE-CTA 的患者。分析了标准线性混合(M_0.6)、70keV 传统 VMI(M70)和 40keV 噪声优化 VMI(M40+)重建。对于 M70 和 M40+数据集,由两位观察者独立确定主观最佳窗口设置(宽度和水平,B-W/L),并使用线性回归计算与主动脉弓衰减相关的优化值(O-W/L)。由另外两位读者评估 W/L 设置之间的图像质量(IQ)的主观评估。采用重复测量方差分析比较 M_0.6、M70 和 M40+之间的 W/L 设置和 IQ 指数。

结果

M70 的 B-W/L 和 O-W/L 分别为 580/210 和 560/200,M40+分别为 1630/570 和 1560/550,均高于标准 DE-CTA 的 W/L 设置(450/100)。关于整体 IQ,M40+的主观评分最高(均 p<0.001)。

结论

必须应用 O-W/L 设置来优化 DE-CTA 的 VMI 重建的主观 IQ。