Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast.

作者信息

Albrecht Moritz H, Scholtz Jan-Erik, Hüsers Kristina, Beeres Martin, Bucher Andreas M, Kaup Moritz, Martin Simon S, Fischer Sebastian, Bodelle Boris, Bauer Ralf W, Lehnert Thomas, Vogl Thomas J, Wichmann Julian L

机构信息

Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.

出版信息

Eur Radiol. 2016 Jun;26(6):1863-70. doi: 10.1007/s00330-015-3970-2. Epub 2015 Sep 3.

Abstract

OBJECTIVES

To compare quantitative image quality parameters in abdominal dual-energy computed tomography angiography (DE-CTA) using an advanced image-based (Mono+) reconstruction algorithm for virtual monoenergetic imaging and standard DE-CTA.

METHODS

Fifty-five patients (36 men; mean age, 64.2 ± 12.7 years) who underwent abdominal DE-CTA were retrospectively included. Mono + images were reconstructed at 40, 50, 60, 70, 80, 90 and 100 keV levels and as standard linearly blended M_0.6 images (60 % 100 kV, 40 % 140 kV). The contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) of the common hepatic (CHA), splenic (SA), superior mesenteric (SMA) and left renal arteries (LRA) were objectively measured.

RESULTS

Mono+ DE-CTA series showed a statistically superior CNR for 40, 50, 60, 70 and 80 keV (P < 0.031) compared to M_0.6 images for all investigated arteries except SMA at 80 keV (P = 0.08). CNR at 40 keV revealed a mean relative increase of 287.7 % compared to linearly blended images among all assessed arteries (P < 0.001). SNR of Mono+ images was consistently significantly higher at 40, 50, 60 and 70 keV compared to M_0.6 for CHA and SA (P < 0.009).

CONCLUSIONS

Compared to linearly blended images, Mono+ reconstructions at low keV levels of abdominal DE-CTA datasets significantly improve quantitative image quality.

KEY POINTS

• Mono+ combines increased attenuation with reduced image noise compared to standard DE-CTA. • Mono+ shows superior contrast-to-noise ratios at low keV compared to linearly-blended images. • Contrast-to-noise ratio in monoenergetic DE-CTA peaks at 40 keV. • Mono+ reconstructions significantly improve quantitative image quality at low keV levels.

摘要

目的

比较使用基于图像的先进(Mono+)重建算法进行虚拟单能成像的腹部双能计算机断层血管造影(DE-CTA)与标准DE-CTA的定量图像质量参数。

方法

回顾性纳入55例接受腹部DE-CTA检查的患者(36例男性;平均年龄64.2±12.7岁)。在40、50、60、70、80、90和100keV水平重建Mono+图像,并作为标准线性混合M_0.6图像(60% 100kV,40% 140kV)。客观测量肝总动脉(CHA)、脾动脉(SA)、肠系膜上动脉(SMA)和左肾动脉(LRA)的对比噪声比(CNR)和信噪比(SNR)。

结果

与M_0.6图像相比,Mono+ DE-CTA系列在40、50、60、70和80keV时显示出统计学上更高的CNR(P<0.031),除80keV时的SMA外,所有研究动脉均如此(P=0.08)。在所有评估动脉中,40keV时的CNR与线性混合图像相比平均相对增加287.7%(P<0.001)。与CHA和SA的M_0.6相比,Mono+图像在40、50、60和70keV时的SNR始终显著更高(P<0.009)。

结论

与线性混合图像相比,腹部DE-CTA数据集在低keV水平的Mono+重建显著提高了定量图像质量。

关键点

•与标准DE-CTA相比,Mono+结合了增加的衰减和降低的图像噪声。•与线性混合图像相比,Mono+在低keV时显示出更高的对比噪声比。•单能DE-CTA中的对比噪声比在40keV时达到峰值。•Mono+重建在低keV水平显著提高了定量图像质量。