Furman J M, Wall C, Kamerer D B
Department of Otolaryngology, University of Pittsburgh, PA
Ann Otol Rhinol Laryngol. 1988 Jul-Aug;97(4 Pt 1):359-64. doi: 10.1177/000348948809700406.
This study compares the results from simultaneous binaural bithermal and alternate binaural bithermal caloric testing of 652 patients with dizziness. Also compared are the abilities of these tests to distinguish between a population of healthy asymptomatic subjects and our patient population. Caloric testing used a closed-loop irrigation system. Eye movements were monitored with electro-oculography. The agreement between the results from the two tests was low. Also, while the simultaneous test was more sensitive, the alternate test was more specific. Using receiver-operator characteristic methodology, it was found that the ability of the alternate test to distinguish between a healthy population and a patient population was superior to that of the simultaneous test. There was no improvement in the ability to distinguish between a healthy population and a patient population when the results from the two tests were combined as compared to using the alternate caloric test alone.
本研究比较了652例头晕患者同时双耳冷热试验和交替双耳冷热试验的结果。还比较了这些测试区分健康无症状受试者群体和患者群体的能力。冷热试验使用闭环冲洗系统。用眼电图监测眼动。两种测试结果之间的一致性较低。此外,虽然同时测试更敏感,但交替测试更具特异性。使用受试者工作特征方法发现,交替测试区分健康人群和患者人群的能力优于同时测试。与单独使用交替冷热试验相比,将两种测试结果结合起来区分健康人群和患者人群的能力没有提高。