Suppr超能文献

磁共振引导激光诱导热疗期间磁共振热成像信号伪影的特征分析

Characterization of Magnetic Resonance Thermal Imaging Signal Artifact During Magnetic Resonance Guided Laser-Induced Thermal Therapy.

作者信息

Munier Sean M, Liang Allison S, Desai Akshay N, James Jose K, Danish Shabbar F

机构信息

Department of Neurosurgery, Rutgers-RWJ Medical School, New Brunswick, New Jersey.

出版信息

Oper Neurosurg (Hagerstown). 2020 Oct 15;19(5):619-624. doi: 10.1093/ons/opaa229.

Abstract

BACKGROUND

Magnetic resonance-guided laser interstitial thermal therapy (MRgLITT) is a minimally invasive procedure that utilizes intraoperative magnetic resonance thermal imaging (MRTI) to generate a thermal damage estimate (TDE) of the ablative area. In select cases, the MRTI contains a signal artifact or defect that distorts the ablative region. No study has attempted to characterize this artifact.

OBJECTIVE

To characterize MRTI signal the artifact in select cases to better understand its potential relevance and impact on the ablation procedure.

METHODS

All ablations were performed using the Visualase magnetic resonance imaging-guided laser ablation system (Medtronic). Patients were included if the MRTI contained signal artifact that distorted the ablative region during the first thermal dose delivered. Ablation artifact was quantified using MATLAB version R2018a (Mathworks Inc, Natick, Massachusetts).

RESULTS

A total of 116 patients undergoing MRgLITT for various surgical indications were examined. MRTI artifact was observed in 37.0% of cases overall. Incidence of artifact was greater at higher powers (P < .001) and with longer ablation times (P = .024), though artifact size did not correlate with laser power or ablation duration.

CONCLUSION

MRTI signal artifact is common during LITT. Higher powers and longer ablation times result in greater incidence of ablation artifact, though artifact size is not correlated with power or duration. Future studies should aim to evaluate effects of artifact on postoperative imaging and, most notably, patient outcomes.

摘要

背景

磁共振引导激光间质热疗(MRgLITT)是一种微创手术,利用术中磁共振热成像(MRTI)生成消融区域的热损伤估计(TDE)。在某些情况下,MRTI包含信号伪影或缺陷,会使消融区域失真。尚无研究试图对这种伪影进行特征描述。

目的

在某些情况下对MRTI信号伪影进行特征描述,以更好地理解其潜在相关性以及对消融手术的影响。

方法

所有消融均使用Visualase磁共振成像引导激光消融系统(美敦力公司)进行。如果在首次热剂量输送期间MRTI包含使消融区域失真的信号伪影,则纳入患者。使用MATLAB R2018a版本(Mathworks公司,马萨诸塞州纳蒂克)对消融伪影进行量化。

结果

共检查了116例因各种手术适应症接受MRgLITT的患者。总体而言,37.0%的病例观察到MRTI伪影。在较高功率(P <.001)和较长消融时间(P =.024)时,伪影发生率更高,尽管伪影大小与激光功率或消融持续时间无关。

结论

在激光间质热疗(LITT)期间,MRTI信号伪影很常见。较高功率和较长消融时间会导致消融伪影发生率更高,尽管伪影大小与功率或持续时间无关。未来的研究应旨在评估伪影对术后成像的影响,最值得注意的是对患者预后的影响。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验