脑脊液脉搏波速度测量:一种新型多带电影相位对比 MRI 序列的体外表征和体内评估。

Cerebrospinal fluid pulse wave velocity measurements: In vitro and in vivo evaluation of a novel multiband cine phase-contrast MRI sequence.

机构信息

Department of Diagnostic and Interventional Radiology, University of Cologne, Faculty of Medicine and University Hospital Cologne, Cologne, Germany.

Department of General Neurosurgery, Center for Neurosurgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Cologne, Germany.

出版信息

Magn Reson Med. 2021 Jan;85(1):197-208. doi: 10.1002/mrm.28430. Epub 2020 Jul 21.

Abstract

PURPOSE

Intracranial and intraspinal compliance are parameters of interest in the diagnosis and prediction of treatment outcome in patients with normal pressure hydrocephalus and other forms of communicating hydrocephalus. A noninvasive method to estimate the spinal cerebrospinal fluid (CSF) pulse wave velocity (PWV) as a measure of compliance was developed using a multiband cine phase-contrast MRI sequence and a foot-to-foot algorithm.

METHODS

We used computational simulations to estimate the accuracy of the MRI acquisition and transit-time algorithm. In vitro measurements were performed to investigate the reproducibility and accuracy of the measurements under controlled conditions. In vivo measurements in 20 healthy subjects and 2 patients with normal pressure hydrocephalus were acquired to show the technical feasibility in a clinical setting.

RESULTS

Simulations showed a mean deviation of the calculated CSF PWV of 3.41% ± 2.68%. In vitro results were in line with theory, showing a square-root relation between PWV and transmural pressure and a good reproducibility with SDs of repeated measurements below 5%. Mean CSF PWV over all healthy subjects was 5.83 ± 3.36 m/s. The CSF PWV measurements in the patients with normal pressure hydrocephalus were distinctly higher before CSF shunt surgery (33.80 ± 6.75 m/s and 31.31 ± 7.82 m/s), with a decrease 5 days after CSF shunt surgery (15.69 ± 3.37 m/s).

CONCLUSION

This study evaluates the feasibility of CSF PWV measurements using a multiband cine phase-contrast MRI sequence. In vitro and in vivo measurements showed that this method is a potential tool for the noninvasive estimation of intraspinal compliance.

摘要

目的

颅内和椎管顺应性是正常压力脑积水和其他交通性脑积水患者诊断和治疗效果预测的关注参数。我们使用多带宽电影相位对比 MRI 序列和足到足算法开发了一种非侵入性方法来估计脊髓脑脊液(CSF)脉搏波速度(PWV)作为顺应性的指标。

方法

我们使用计算模拟来估计 MRI 采集和渡越时间算法的准确性。在体外测量中,我们在受控条件下进行了测量的可重复性和准确性的研究。在 20 名健康受试者和 2 名正常压力脑积水患者中进行了体内测量,以在临床环境中展示技术的可行性。

结果

模拟显示计算的 CSF PWV 的平均偏差为 3.41%±2.68%。体外结果与理论相符,显示 PWV 与跨壁压力之间存在平方根关系,并且具有良好的可重复性,重复测量的标准差低于 5%。所有健康受试者的平均 CSF PWV 为 5.83±3.36 m/s。正常压力脑积水患者在 CSF 分流手术前的 CSF PWV 测量值明显较高(33.80±6.75 m/s 和 31.31±7.82 m/s),CSF 分流手术后 5 天降低(15.69±3.37 m/s)。

结论

本研究评估了使用多带宽电影相位对比 MRI 序列测量 CSF PWV 的可行性。体外和体内测量表明,该方法是一种用于无创估计椎管顺应性的潜在工具。