基于 4D 流 MRI 的插值时间优化的主动脉脉搏波速度估计。

Interpolation time-optimized aortic pulse wave velocity estimation by 4D flow MRI.

机构信息

Daegu-Gyeongbuk Medical Innovation Foundation, Medical Device Development Center, Daegu, 41061, South Korea.

Institute of Medical Devices, Kangwon National University, Chuncheon, South Korea.

出版信息

Sci Rep. 2023 Sep 30;13(1):16484. doi: 10.1038/s41598-023-43799-z.

Abstract

Four-dimensional flow magnetic resonance imaging-based pulse wave velocity (4D flow PWV) estimation is a promising tool for measuring regional aortic stiffness for non-invasive cardiovascular disease screening. However, the effect of variations in the shape of flow waveforms on 4D flow PWV measurements remains unclear. In this study, 4D flow PWV values were compared using cross-correlation algorithm with different interpolation times (iTs) based on flow rate and beat frequency. A critical iT (iT) was proposed from in vitro study using flexible and stiff phantom models to simultaneously achieve a low difference and a low computation time. In vivo 4D flow PWV values from six healthy volunteers were also compared between iT and the conventionally used interpolation time of 1 ms (iT). The results indicated that iT reduced the mean difference of in vitro 4D flow PWV values by 19%, compared to iT. In addition, iT measured in vivo 4D flow PWV, showing differences similar to those obtained with iT. A difference estimation model was proposed to retrospectively estimate potential differences of 4D flow PWV using known values of PWV and the used iT. This study would be helpful for understanding the differences of PWV generated by physiological changes and time step of obtained flow waveforms.

摘要

基于四维流磁共振成像的脉搏波速度(4D flow PWV)估计是一种很有前途的工具,可用于测量区域性主动脉僵硬度,进行非侵入性心血管疾病筛查。然而,血流波形形状变化对 4D flow PWV 测量的影响尚不清楚。在这项研究中,基于流率和心跳频率,使用互相关算法比较了不同内插时间(iT)的 4D flow PWV 值。通过使用柔性和刚性体模的体外研究提出了一个临界 iT(iT),以便同时实现低差值和低计算时间。还比较了六个健康志愿者的体内 4D flow PWV 值,比较了 iT 与传统的 1 ms 内插时间(iT)。结果表明,与 iT 相比,iT 降低了体外 4D flow PWV 值的平均差值 19%。此外,iT 测量了体内 4D flow PWV,其差异与 iT 相似。提出了一种差异估计模型,用于使用已知的 PWV 值和使用的 iT 来回顾性估计 4D flow PWV 的潜在差异。这项研究有助于理解由生理变化和获得的血流波形时间步长引起的 PWV 差异。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4c6f/10542805/87f207ddbb13/41598_2023_43799_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验