Browne Jacinta E, Tiegs-Heiden Christin A, Lehman Vance T, Long Zaiyang, Hangiandreou Nicholas J, Watson Robert E, Hesley Gina K, Gorny Krzysztof R
Department of Radiology, Mayo Clinic, Rochester, MN 55905.
Mayo Clin Proc Innov Qual Outcomes. 2020 Aug 5;4(4):464-468. doi: 10.1016/j.mayocpiqo.2020.04.010. eCollection 2020 Aug.
To provide an initial report that patients with magnetic resonance imaging (MRI) non-conditional cardiac implanted electronic device (CIED) can undergo state-of-the-art magnetic resonance imaging-guided focused (MRgFUS) ablation procedures with careful planning and integration of the procedure into an established CIED MRI practice.
We describe an MRgFUS ablation treatment of lumbar facet joints in a patient with an MRI non-conditional CIED (pacemaker), completed in accordance with our institutional CIED/MRI practice guidelines.
A risk-benefit analysis by a coordinated multidisciplinary team before this treatment was performed to account for the risks associated with the MRI non-conditional pacemaker in the context of the MRgFUS procedure.
The patient had no adverse cardiac event during or following this procedure.
提供一份初步报告,表明磁共振成像(MRI)非条件性心脏植入式电子设备(CIED)患者在经过精心规划并将该手术纳入既定的CIED MRI操作流程后,能够接受先进的磁共振成像引导聚焦超声(MRgFUS)消融手术。
我们描述了一名患有MRI非条件性CIED(起搏器)患者的腰椎小关节MRgFUS消融治疗,该治疗按照我们机构的CIED/MRI操作指南完成。
在进行该治疗前,由多学科协作团队进行了风险效益分析,以考虑在MRgFUS手术背景下与MRI非条件性起搏器相关的风险。
该患者在手术期间及术后未发生不良心脏事件。