Magnetic resonance imaging in patients with cardiac implanted electronic devices: focus on contraindications to magnetic resonance imaging protocols.

作者信息

Horwood Laura, Attili Anil, Luba Frank, Ibrahim El-Sayed H, Parmar Hemant, Stojanovska Jadranka, Gadoth-Goodman Sharon, Fette Carey, Oral Hakan, Bogun Frank

机构信息

Division of Cardiovascular Medicine, University of Michigan, 1500 E. Medical Center Dr., SPC 5853, Ann Arbor, MI 48109-5853, USA.

Department of Radiology, University of Michigan, Ann Arbor, MI, USA.

出版信息

Europace. 2017 May 1;19(5):812-817. doi: 10.1093/europace/euw122.

Abstract

AIMS

Magnetic resonance imaging (MRI) has been reported to be safe in patients with cardiac implantable electronic devices (CIED) provided a specific protocol is followed. The objective of this study was to assess whether this is also true for patients excluded from published protocols.

METHODS AND RESULTS

A total of 160 MRIs were obtained in 142 consecutive patients with CIEDs [106 patients had an implantable cardioverter defibrillator (ICD) and 36 had a pacemaker implanted] using an adapted, pre-specified protocol. A cardiac MRI was performed in 95 patients, and a spinal/brain MRI was performed in 47 patients. Forty-six patients (32%) had either abandoned leads (n = 10), and/or were pacemaker dependent with an implanted ICD (n = 19), had recently implanted CIEDs (n = 1), and/or had a CIED device with battery depletion (n = 2), and/or a component of the CIED was recalled or on advisory (n = 32). No major complications occurred. Some device parameters changed slightly, but significantly, right after or at 1-week post-MRI without requiring any reprogramming. In one patient with an ICD on advisory, the pacing rate changed inexplicably during one of his two MRIs from 90 to 50 b.p.m.

CONCLUSION

Using a pre-specified protocol, cardiac and non-cardiac MRIs were performed in CIED patients with pacemaker dependency, abandoned leads, or depleted batteries without occurrence of major adverse events. Patients with devices on advisory need to be monitored carefully during MRI, especially if they are pacemaker dependent.

摘要

目的

据报道,只要遵循特定方案,心脏植入式电子设备(CIED)患者进行磁共振成像(MRI)是安全的。本研究的目的是评估对于被已发表方案排除在外的患者情况是否也是如此。

方法与结果

采用预先指定的调整方案,对142例连续的CIED患者共进行了160次MRI检查[106例患者植入了植入式心律转复除颤器(ICD),36例植入了起搏器]。95例患者进行了心脏MRI检查,47例患者进行了脊柱/脑部MRI检查。46例患者(32%)有废弃导线(n = 10),和/或依赖起搏器且植入了ICD(n = 19),近期植入了CIED(n = 1),和/或CIED设备存在电池耗尽(n = 2),和/或CIED的某个部件被召回或处于咨询状态(n = 32)。未发生重大并发症。一些设备参数在MRI检查后即刻或1周时略有但显著变化,无需任何重新编程。在1例处于咨询状态的ICD患者中,其两次MRI检查中的一次期间起搏频率莫名从90次/分钟变为50次/分钟。

结论

采用预先指定的方案,对有起搏器依赖、废弃导线或电池耗尽的CIED患者进行了心脏和非心脏MRI检查,未发生重大不良事件。处于咨询状态的设备患者在MRI检查期间需要仔细监测,尤其是那些依赖起搏器的患者。